GE Healthcare, LLC: Discovery NM/CT 670 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:

Recall number
Z-1259-2019
Recalling firm
GE Healthcare, LLC
Product
Discovery NM/CT 670 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:
Status
Terminated
Initiated
2019-03-20
Posted
not on file
Terminated
2020-10-05
Reason
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Root cause
Device Design
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
11 units
Distribution
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Lot, serial or model codes
Mfg. Lot or Serial # System ID Model, Catalogue or Code # DRGX56002 SHIPPED (GON 4683652) H3100BT DDRX34058 909558D670NM H3100AN DDRX34056 650723D670 H3100AN DDRX34053 SHIPPED (AO 1614951) H3100AN DRGX56005 614CVA6702 H3100BT DRGX56017 110002NU06 H3100BT DRGX56004 403343D670A H3100BT DRGX56018 X7410502 H3100BT DRGX56009 083026206531618 H3100BT DRGX56012 STD01300 H3100BT DDRX34055 00362NUC04 H3100AN
510(k) numbers
["K093514"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register