Skip to the content

FDA Medical Device Recalls (openFDA)

Focalyx Technologies, LLC.: Focalyx Fusion

Recall number
Z-1243-2026
Recalling firm
Focalyx Technologies, LLC.
Product
Focalyx Fusion
Status
Open, Classified
Initiated
2025-12-23
Posted
2026-02-03
Terminated
not on file
Reason
Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.
Root cause
Software change control
Action
On 12/23/2025, correction notices started being emailed to customers informing them of the following: You will be contacted to schedule installation of the new version. An IFU will be provided at time of installation. Complete and return the response form via email to corporate@focalyx.com If there are any questions, contact the firm at: focalyxtechnologies.com/contact, support@focalyxtechnologies.com, (888) 636 2259, (Sel. 3 Technical Support)
Quantity
18
Distribution
Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.
Lot, serial or model codes
UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324
510(k) numbers
["K170250"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3016567400
Firm city
Hialeah
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register