Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics, Inc.: VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)

Recall number
Z-1230-2024
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Status
Open, Classified
Initiated
2024-01-26
Posted
2024-03-01
Terminated
not on file
Reason
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
Root cause
Software Design Change
Action
On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules. The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Quantity
15 US; 43 OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Lot, serial or model codes
VITROS System Software Versions 3.8.0 or 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers: US: J46000208 J46000209 J46000240 J46000497 J46000864 J46000865 J46000898 J46001030 J46001064 J46001125 J46001132 J46001293 J46001462 J46001703 J46001704 OUS: J46000150 J46000933 J46000162 J46001029 J46000190 J46001041 J46000247 J46001042 J46000296 J46001087 J46000393 J46001119 J46000410 J46001186 J46000478 J46001187 J46000499 J46001210 J46000531 J46001217 J46000564 J46001289 J46000617 J46001541 J46000772 J46001565 J46000788 J46001571 J46000800 J46001609 J46000801 J46001629 J46000812 J46001639 J46000821 J46001645 J46000855 J46001680 J46000859 J46000948 J46000861 J46000983 J46000910
510(k) numbers
K031924
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register