FDA Medical Device Recalls (openFDA)
Inova Diagnostics Incorporated: QUANTA-Lyser 240 IFA, Part #GS0242. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, dispensing slide mounting medium in the preparation for microscope analysis, and photometric measurement of microwell plates.
- Recall number
- Z-1221-2018
- Recalling firm
- Inova Diagnostics Incorporated
- Product
- QUANTA-Lyser 240 IFA, Part #GS0242. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, dispensing slide mounting medium in the preparation for microscope analysis, and photometric measurement of microwell plates.
- Status
- Terminated
- Initiated
- 2018-01-26
- Posted
- not on file
- Terminated
- 2018-05-22
- Reason
- The device power supply was not manufactured according to specification which can result in the loss of the protective grounding of the instrument.
- Root cause
- Nonconforming Material/Component
- Action
- The letters were issued via email on January 26, 2018, informing the consignee of the issue and the actions to be taken. The response forms were requested to be completed and returned by February 23, 2018, indicating receipt of the notification and confirmation that the power supply has been replaced. The firm reported on February 23, 2018 that responses have been received from both consignees. For further questions, please call (858) 586-9900 x 1381.
- Quantity
- 1 instrument
- Distribution
- There was no U.S. distribution. Distribution was made to Panama and United Kingdom.
- Lot, serial or model codes
- Serial number 4314
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2026994
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28