Miach Orthopaedics: BEAR Implant. Model Number: 1000.
- Recall number
- Z-1220-2026
- Recalling firm
- Miach Orthopaedics
- Product
- BEAR Implant. Model Number: 1000.
- Status
- Open, Classified
- Initiated
- 2025-12-31
- Posted
- 2026-01-29
- Terminated
- not on file
- Reason
- Incorrect expiration date on label that extends beyond the product's documented/approved shelf life. With an incorrect expiration date, the UDI will also be incorrect.
- Root cause
- Process control
- Action
- Notification to Sales Representatives and consignees began on about 12/31/2025 via email and letter titled URGENT MEDICAL DEVICE RECALL NOTIFICATION. Consignees were instructed to identify and quarantine all unused inventory of the affected lot, discontinue distribution of the affected lot, return all unused affected product, and complete and return the provided Recall Acknowledgement Form. Additionally, consignees were requested to notify customers if the affected lot was further distributed.
- Quantity
- 284 units
- Distribution
- US Nationwide distribution in the states of AK, CA, CO, CT, FL, GA, IL, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, TX, VA, WA, WI.
- Lot, serial or model codes
- Model Number: 1000. UDI Number: (01)00860002987804(17)260228(10)7011098. Lot Number: 7011098. Expiration Date: 02/28/2026.
- 510(k) numbers
- ["K243578"]
- PMA numbers
- not on file
- Product code
- QNI
- Firm FEI
- 3020652309
- Firm city
- Westborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28