Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: 1.5mm System, Cranio-maxillofacial Bone Screws 1.5 x 4mm HT SD X-DR Screw, Bone Screw

Recall number
Z-1220-2017
Recalling firm
Zimmer Biomet, Inc.
Product
1.5mm System, Cranio-maxillofacial Bone Screws 1.5 x 4mm HT SD X-DR Screw, Bone Screw
Status
Terminated
Initiated
2017-01-10
Posted
not on file
Terminated
2018-03-26
Reason
The identification of the devices was non-conforming with off-centered slots and the no-go gage starting in the pilot hole.
Root cause
Under Investigation by firm
Action
Zimmer Biomet initiated a voluntary recall of 1.5mm System due to identification of non-conformance with off-centered slots and the no-go gage starting in the pilot hole. One distributor with affected product outside of Zimmer Biomet control was notified via email and asked to return the affected product. This communication was sent on via e-mail on January 10, 2017. All product has been returned.There were 89 units in Zimmer Biomet control at the time the removal was initiated, however all 100 are now in Zimmer Biomet control. Customer Service Zimmer Biomet, Inc. P.O. Box 708 Warsaw, IN 46581-0708 Phone: 800-348-2759
Quantity
11
Distribution
Domestic: None VA/DOD: None Foreign: Japan
Lot, serial or model codes
95-6104- Item No.  530430- Lot 1.5x4mm HT SD X-DR Screw
510(k) numbers
["K121589"]
PMA numbers
not on file
Product code
HWC
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register