Alpha Omega Engineering: Neuro Omega System, incorporating HaGuide software. For neurological and neurosurgical use. To monitor, record and display bioelectric signals, and stimulate nerves and brain tissue.

Recall number
Z-1216-2019
Recalling firm
Alpha Omega Engineering
Product
Neuro Omega System, incorporating HaGuide software. For neurological and neurosurgical use. To monitor, record and display bioelectric signals, and stimulate nerves and brain tissue.
Status
Terminated
Initiated
2018-09-04
Posted
2019-05-06
Terminated
2020-11-03
Reason
If cables are improperly connected, current may reach high charge density, causing tissue harm.
Root cause
Device Design
Action
The firm originally notified customers of the issue via email on September 4, 2018. Also, another updated notification Urgent Field Safety Notice, Neuro Omega system, DATE March 4, 2019, Field Advisory Note [FAN] Attention: Details on affected devices: Neuro Omega system with: Catalogue number: (Neuro Omega) NOM-000000-00 + (Drive) 750-000020-00 + (EMG) 750-000010-00 + (EEG) 750-000010-12 Software version number: Any, clarifying that the issue only occurred in a triple-fault condition.
Quantity
29
Distribution
US nationwide distribution: AL, AZ, CA, CO, GA, IL, LA, MD, MA, MI, MN, OH, PA, RI, TX, and VA
Lot, serial or model codes
Serial numbers: 113, 115, 116, 118, 123, 121, 125, 129, 130, 137, 138, 141, 140, 146, 149, 147, 156/04, 159/04, 163/04, 161/04, 167, 175/06, 176/06, 177/06, 180/06, 182/06, 185/06, 188/06, and 189.
510(k) numbers
["K123796","K171581"]
PMA numbers
not on file
Product code
GZL
Firm FEI
3001610756
Firm city
Nazerat Illit
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register