DiaSorin Molecular LLC: Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.

Recall number
Z-1210-2023
Recalling firm
DiaSorin Molecular LLC
Product
Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.
Status
Open, Classified
Initiated
2023-01-30
Posted
2023-03-03
Terminated
not on file
Reason
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Root cause
Device Design
Action
On 01/30/2023, correction notices were distributed to customers via in-person site visits. Customers were asked to do the following: 1) Discontinue use of affected devices using affected assay definitions. 2) Customers can contact the recalling firm to replace affected kits. The firm can also provide assistance with inventory reconciliation/destruction, and if needed, assistance with potential instrument/lab decontamination. 3) Any sample result with an error code (EC510, EC515, RNA IC, Insufficient Specimen Volume) should be retested following the Instructions for Use. 4) If leakage is suspected, the spillage should be removed and the instrument surface should be decontaminated. Ensure the practice of Good Laboratory Procedures including regular cleaning. 5) This recall notice should be shared with anyone who needs to be aware within your organization or shared with any organization where the potentially affected products have been transferred. Notify others within your network that may have remaining inventory of those kits. 5) Complete and return the acknowledgement and receipt form and return to the recalling firm's Technical Service department at technicalinfo.molecular@diasorin.com Product in the field with the previous high spin assay definitions are being recalled and replaced with current products with the current reduced spin assay definitions. Customers with questions are encouraged to email (see above) or call Technical Service at (800) 838-4548, option 3, Monday - Friday, 7:00am to 5:00pm (Pacific Time)
Quantity
4142
Distribution
US Nationwide distribution including in the states of NJ, ME, ND, IL, MI, FL, GA, AZ, MA, CA, MO, SD, MN, PA, OH, VT, NE, WI, CO, DC, VA, TX, NH, IN, IA, MD, NC, AL, NY, KY, WV, RI, KS, NM, OK, WA, SC, CT, TN, NV, LA.
Lot, serial or model codes
UDI-DI: 20816101026962. Product REF/Lot/Reaction Mix REF/Lot: MOL2655/13648N/MOL2656/13649N, MOL2655/13882N/MOL2656/15164N, MOL2655/15279N/MOL2656/15280N
510(k) numbers
["K201505"]
PMA numbers
not on file
Product code
OCC
Firm FEI
2023365
Firm city
Cypress
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register