Skip to the content

FDA Medical Device Recalls (openFDA)

DiaDexus, Inc: PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit, assembled in a kit box with an outer label, Catalog numbers 90108 and 10-0108, manufactured by diaDexus inc., South San Francisco, CA. Intended to establish points of reference that are used in the determination of values in the determination of Lp-PLA2 by the PLAC Test Reagent Kit.

Recall number
Z-1205-2009
Recalling firm
DiaDexus, Inc
Product
PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit, assembled in a kit box with an outer label, Catalog numbers 90108 and 10-0108, manufactured by diaDexus inc., South San Francisco, CA. Intended to establish points of reference that are used in the determination of values in the determination of Lp-PLA2 by the PLAC Test Reagent Kit.
Status
Terminated
Initiated
2009-02-09
Posted
2009-04-23
Terminated
2010-01-05
Reason
Premature Expiration-- the product is projected to fall short of expected shelf life, and may result in inaccurate results.
Root cause
Process control
Action
Distributors and Direct accounts were notified by phone followed by Notification letter and certificate of destruction via FedEx. International customers were notified via e-mail with PDF copies of notices, to be followed by hard copies via FedEx. Customers are requested to record the number of kits (full and partial ) they had on the Certificate of Destruction and to dispose of that number of kits with two people signing the form, and then to fax that form to diaDexus.
Quantity
105 kits
Distribution
Worldwide Distribution
Lot, serial or model codes
lot number 048091, expiration date 2009-09.
510(k) numbers
["K072599"]
PMA numbers
not on file
Product code
NOE
Firm FEI
3003643666
Firm city
South San Francisco
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register