Skip to the content

FDA Medical Device Recalls (openFDA)

DiaDexus, Inc: diaDexus Inc. PLAC Test Reagent Kit, Turbidimetric Immunoassay for the Quantitative Determination of Lp-PLA2 in Human Plasma or Serum, liquid ready-to-use two-reagent kit assembled in a kit box, manufactured by diaDexus Inc, South San Francisco, CA 94080, Catalog number 90110 Turbidimetric Immunoassay for the Quantitative Determination of Lp-PLA2 in Human Plasma or Serum on automated clinical chemistry analyzers, to be used in conjunction with clinical evaluations and patient risk assessment as an aide in predicting risk for coronary heart disease, and ischemic stroke associated with atherosclerosis.

Recall number
Z-1204-2009
Recalling firm
DiaDexus, Inc
Product
diaDexus Inc. PLAC Test Reagent Kit, Turbidimetric Immunoassay for the Quantitative Determination of Lp-PLA2 in Human Plasma or Serum, liquid ready-to-use two-reagent kit assembled in a kit box, manufactured by diaDexus Inc, South San Francisco, CA 94080, Catalog number 90110 Turbidimetric Immunoassay for the Quantitative Determination of Lp-PLA2 in Human Plasma or Serum on automated clinical chemistry analyzers, to be used in conjunction with clinical evaluations and patient risk assessment as an aide in predicting risk for coronary heart disease, and ischemic stroke associated with atherosclerosis.
Status
Terminated
Initiated
2008-11-04
Posted
2009-04-01
Terminated
2009-12-21
Reason
Storage equipment failure at distributor site due to Hurricane Ike caused product to experience a temperature variation and go out of range. The integrity of the product could not be assured.
Root cause
Other
Action
The firm and its distributor contacted the consignees via phone on 11/04/08. The firm informed them of the issue and inquired about the products' status. An Urgent: Device Recall letter was issued on 11/04/08 via FedEx. The letter requested that use of the kit be immediately discontinued and returned promptly. A phone call was made to follow up requesting return of the product.
Quantity
27 units
Distribution
Nationwide Distribution
Lot, serial or model codes
sub lot of lot 088061, expiration date 2008-12
510(k) numbers
not on file
PMA numbers
not on file
Product code
NOE
Firm FEI
3003643666
Firm city
South San Francisco
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register