NxStage MDS Corporation: NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.

Recall number
Z-1191-2024
Recalling firm
NxStage MDS Corporation
Product
NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.
Status
Open, Classified
Initiated
2024-01-29
Posted
2024-02-27
Terminated
not on file
Reason
Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.
Root cause
Component design/selection
Action
The firm began notifying affected customers on January 29, 2024. Customers were informed that the issue may occur when the needleless access port located next to the filter has been accessed and there is a buildup of pressure in the circuit. Customers were instructed to inspect their inventory for affected product and quarantine recalled product. Customers should not use affected product and should ensure all users of the CAR-535 are aware of the notice. Customers should contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243) to arrange for the return of all affected products. This number may also be contacted if you have any questions or comments.
Quantity
693
Distribution
US domestic distribution to Dialysis Centers located in AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI. No international distribution.
Lot, serial or model codes
UDI-DI M535CAR5350; Package DI M535CAR5355 Lots 20878001, 30978013, 30978032, 31178011, 31178019
510(k) numbers
["K183158"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3003464075
Firm city
Lawrence
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register