FDA Medical Device Recalls (openFDA)

Getinge Group Logistics America, LLC: Anesthesia gas-machine - Product Usage: intended for use in administering anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with limited ability to breathe.

Recall number
Z-1181-2021
Recalling firm
Getinge Group Logistics America, LLC
Product
Anesthesia gas-machine - Product Usage: intended for use in administering anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with limited ability to breathe.
Status
Terminated
Initiated
2021-01-29
Posted
not on file
Terminated
2023-09-28
Reason
Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.
Root cause
Device Design
Action
On 2/9/2021 Maquet Critical Care AB / Getinge issued an Urgent Medical Device Correction letter via FedEx for the Flow i anesthesia machine due to the potential for high FiCO2 values and/or a System Check Out failure due to an incomplete docking of the CO2 absorber to the patient cassette. The patients re-breathing without CO2 reduction could occur if the absorber is not sufficiently docked.
Quantity
144 units
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, MA, MI, MO, NJ, NY, OH, OR, PA, TX, VA, WA, WV.
Lot, serial or model codes
UDI Code: Flow i C20 Anesthesia System 6677200 07325710001349, Flow i C30 Anesthesia System 6677300 07325710001349, Flow i C40 Anesthesia System 6677400 07325710001318, Flow c Anesthesia System 6887700 07325710009765 Serial Number: 1172, 1546, 2095, 2719 2724, 5427 5429, 6132, 6819, 1202, 1643 1657, 2186, 2729 2733, 5540 5542, 6133, 6820, 1204,1786, 2553 2555, 2735 2747, 5589, 6525, 6821, 1224 1226, 1787, 2557 2562, 2851 2856, 5590, 6562, 6822, 1277, 1860, 2613, 2858 2869, 5812 5814, 6736, 6823, 1280 1282, 1861, 2714, 4548 4550, 5895 6737, 6824, 1545, 2021 2024, 2716, 5198 5203, 6077, 6818, 6940, 1058, 1059, 1060, 1167, 1169, 1201, 2034, 5204 5211, 7185, 4003.
510(k) numbers
K191027
PMA numbers
not on file
Product code
BSZ
Firm FEI
3025504317
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source