MRP, LLC dba Aquabiliti: AQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, intermediate cartons containing 100 syringes, shipping cases containing 400 syringes, Rx, Sterile.

Recall number
Z-1174-2024
Recalling firm
MRP, LLC dba Aquabiliti
Product
AQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, intermediate cartons containing 100 syringes, shipping cases containing 400 syringes, Rx, Sterile.
Status
Open, Classified
Initiated
2023-12-28
Posted
2024-02-22
Terminated
not on file
Reason
The saline flush syringes may fail to meet USP criteria for stated sodium chloride content after 12 months.
Root cause
Under Investigation by firm
Action
The recalling firm issued a letter dated 12/28/2023 via email. The letter explained the reason for recall, risk to health, and actions to be taken. The actions included the consignee examine their inventory and discontinue distribution and use immediately. All products should be segregated. A Shipment Summary and Acknowledgment Form was enclosed for completion and return to indicate the consignee has read and understands the recall instructions, identified and quarantined the affected product in inventory, and that the consignee has notified their customers who were shipped the product. It goes onto list the specific lot numbers and quantity of affected product sent to the consignee and asks for the quantity of product consumed and the quantity available for return. The letter was also issued via certified mail on 1/4/2024. The firm issued a corrected letter dated 1/8/2024 to correct the table listing lot numbers on Page 2 of the letter. Those lot numbers were asterisked for ease of identification by the consignee. The lot numbers listed on the Shipment Summary and Acknowledgment Form were correct in the original letter.
Quantity
3,080,000 syringes
Distribution
US Nationwide distribution in the states CO, FL, NC, NY, OH, PA, TN, TX, and VA.
Lot, serial or model codes
REF 2T0805 - Lot numbers KH04910, exp. 1/11/2024; KH04930, exp. 3/13/2024; KH04954, exp. 5/22/2024; KH04975, exp. 8/2/2024; KH04990, exp. 9/6/2024; KH05020, exp. 10/23/2024; KH05022, exp. 10/24/2024; KH05043, exp. 11/13/2024; KH05066, exp. 12/13/2024; KH05068, exp. 12/14/2024; KH05070, exp. 12/18/2024; KH05073, exp. 12/20/2024; KH05134, exp. 6/25/2025; KH05135, exp. 6/27/2025; and KH05152, exp. 10/10/2025; UDI-DI 10859809005013. REF 2T0807 - Lot numbers KH04909, exp. 1/10/2024; KH04944, exp. 4/24/2024; KH04986, exp. 8/30/2024; KH04988, exp. 9/5/2024; KH05024, exp. 10/25/2024; KH05026, exp. 10/26/2024; and KH05071, exp. 12/19/2024; UDI-DI 10859809005037.
510(k) numbers
K113555
PMA numbers
not on file
Product code
NGT
Firm FEI
1000151124
Firm city
Nashville
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register