Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Sensis / Sensis Vibe systems with software version VD12 and equipped with HP Flex Pro-C PC. Diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies Model Numbers: Sensis Vibe Hemo 11007641 Sensis Vibe Combo 11007642 Sensis 10764561 Postprocessing Workstation 6648161 SIS Server 6648153

Recall number
Z-1173-2021
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Sensis / Sensis Vibe systems with software version VD12 and equipped with HP Flex Pro-C PC. Diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies Model Numbers: Sensis Vibe Hemo 11007641 Sensis Vibe Combo 11007642 Sensis 10764561 Postprocessing Workstation 6648161 SIS Server 6648153
Status
Terminated
Initiated
2021-01-27
Posted
not on file
Terminated
2021-10-25
Reason
System may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system
Root cause
Software design
Action
Siemens issued Urgent Medical Device Correction via Customer Safety Advisory Notice AX010/21/S dated 1/27/21distributed to all affected customers. Additionally, a solution to eliminate the root cause of this problem is being distributed via Update Instruction AX009/21/S to all affected customers. Letter states reason for recall, health risk and action to take: The freeze is a total lock-up of the system and requires a power cycle of the PC to recover. Please make sure that an alternative system can be used to continue treatment in such cases.Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Quantity
37 US
Distribution
US Nationwide distribution in the states of AK, AR, CA, GU, IA, IL, MI, MO, NC, NE, OR, PA, PA, PR, SD, TN, TX, WA.
Lot, serial or model codes
Software version VD12 and equipped with HP Flex Pro-C PC Serial Numbers:  10704 18900 18901 103008 103014 103019 103022 103039 103040 103042 103047 103048 103053 103079 103080 103082 103086 103088 103089 103090 103115 103116 103117 103119 103123 103124 103130 103131 103137 103144 103145 103148 103152 103160 103170 103171 123004
510(k) numbers
["K150493"]
PMA numbers
not on file
Product code
DQK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register