Biomet, Inc.: Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction

Recall number
Z-1173-2018
Recalling firm
Biomet, Inc.
Product
Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction
Status
Terminated
Initiated
2018-02-28
Posted
2018-03-23
Terminated
2020-08-11
Reason
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.
Root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE RECALL letter will be issued to customers titled "Affected Product: Comprehensive Segmental Revision System (SRS)" instructing customers to do the following: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete the attached Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. ***Updated 3/31/2020*** On March 31, 2020, Zimmer Biomet expanded the recall to include two additional lots: Lot 434480 (for Item No. 211253) and Lot 434470 (for Item No. 211251). Customer instructions remain the same.
Quantity
634
Distribution
Distribution US Nationwide
Lot, serial or model codes
Model numbers: 211250 211251 211252 211253 211254 211255  Lot codes: 095580 119120 130020 142530 155910 162450 162460 241160 257860 304290 305860 309730 337610 389790 400730 419380 430570 443670 545430 546260 548290 550390 586120 606830 611540 636370 636470 688540 704020 714760 793670 806070 829520 829530 881230 950270 970920 019600 110780 130040 157180 185620 203520 241180 278580 330870 364380 424550 430580 457210 544340 583360 636520 706690 714810 810260 853250 881330 883830 891880 923660 923680 991400 033700 053800 062210 110790 160770 185630 241190 247810 257870 347080 583370 611530 636680 638640 674650 688550 775020 829540 848530 889330 950310 976390 988380 991410 028450 028460 053810 062220 098020 155900 160780 231490 241230 251120 251920 257880 281140 330880 347100 367850 400390 450780 493220 494170 527200 546290 548460 548470 548580 586130 611520 636710 665880 704030 761210 829550 853260 889810 914750 923670 923690 925920 927420 937130 991320 062830 119130 130070 185590 196660 201270 241290 247280 247290 257890 338480 347110 350590 366310 554450 636760 648150 704040 714830 750840 767880 793710 829560 878430 927430 948090 952960 970930 976410 040470 192770 241300 258540 288580 293740 320100 340150 347700 364660 419900 442840 462480 462490 569770 611040 611930 688560 770690 780490 815100 818160 829690 867530 891270 923980 924160 927440 201270 848530 976410 033700 350590 554450  ***Lots added 3/31/2020*** Item Number 211253 - Lot Number 434480 - UDI (1)00880304850330(17)280423(10)434480  Item Number 211251 - Lot Number 434470 - UDI (1)00880304850316(17)280507(10)434470
510(k) numbers
["K111746"]
PMA numbers
not on file
Product code
KWT
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register