FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.
- Recall number
- Z-1170-2021
- Recalling firm
- Boston Scientific Corporation
- Product
- Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.
- Status
- Open, Classified
- Initiated
- 2021-01-19
- Posted
- not on file
- Terminated
- not on file
- Reason
- Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
- Root cause
- Process change control
- Action
- On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
- Quantity
- 335,788 units total (Domestic: 222,882, Foreign: 112,906)
- Distribution
- US nationwide distribution.
- Lot, serial or model codes
- Outer box GTIN: 08714729283904 Inner pouch GTIN: 08714729747598 Lot/Batch#: 24818278, 24906823, 25042455 Outer box GTIN: 08714729283928 Inner pouch GTIN: 08714729747611 Lot/Batch#: 24685766, 24817145, 24899710, 24739786, 25006735, 24861901 Outer box GTIN: 08714729158110 Inner pouch GTIN: 08714729748069 Lot/Batch#: 24685764, 24890051, 24824132 Outer box GTIN: 08714729268802 Inner pouch GTIN: 08714729748069 Lot/Batch#: 25023345 Outer box GTIN: 08714729158141 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24699718, 24939626, 24861739 Outer box GTIN: 08714729268819 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24824134, 25006729, 24709534 Outer box GTIN: 08714729501664 Inner pouch GTIN: 08714729748083 Lot/Batch#: 25023203, 24697822, 24739587, 24942292, 24909214, 24824136 Outer box GTIN: 08714729158158 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24856636, 24927642, 24699218 Outer box GTIN: 08714729268826 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24835709, 24818605, 25023343 Outer box GTIN: 08714729501671 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24909216, 24758832, 24872119, 24994171, 25048397, 2475126, 24929431, 24995438
- 510(k) numbers
- ["K131700"]
- PMA numbers
- not on file
- Product code
- FDI
- Firm FEI
- 3005099803
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as