Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.

Recall number
Z-1170-2021
Recalling firm
Boston Scientific Corporation
Product
Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.
Status
Open, Classified
Initiated
2021-01-19
Posted
not on file
Terminated
not on file
Reason
Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Root cause
Process change control
Action
On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
Quantity
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Distribution
US nationwide distribution.
Lot, serial or model codes
Outer box GTIN: 08714729283904 Inner pouch GTIN: 08714729747598 Lot/Batch#: 24818278, 24906823, 25042455  Outer box GTIN: 08714729283928 Inner pouch GTIN: 08714729747611 Lot/Batch#: 24685766, 24817145, 24899710, 24739786, 25006735, 24861901  Outer box GTIN: 08714729158110 Inner pouch GTIN: 08714729748069 Lot/Batch#: 24685764, 24890051, 24824132  Outer box GTIN: 08714729268802 Inner pouch GTIN: 08714729748069 Lot/Batch#: 25023345  Outer box GTIN: 08714729158141 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24699718, 24939626, 24861739  Outer box GTIN: 08714729268819 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24824134, 25006729, 24709534  Outer box GTIN: 08714729501664 Inner pouch GTIN: 08714729748083 Lot/Batch#: 25023203, 24697822, 24739587, 24942292, 24909214, 24824136  Outer box GTIN: 08714729158158 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24856636, 24927642, 24699218  Outer box GTIN: 08714729268826 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24835709, 24818605, 25023343  Outer box GTIN: 08714729501671 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24909216, 24758832, 24872119, 24994171, 25048397, 2475126, 24929431, 24995438
510(k) numbers
["K131700"]
PMA numbers
not on file
Product code
FDI
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register