FDA Medical Device Recalls (openFDA)
St. Jude Medical, Inc.: Proclaim DRG Implantable Pulse Generator, Model Number 3664
- Recall number
- Z-1170-2018
- Recalling firm
- St. Jude Medical, Inc.
- Product
- Proclaim DRG Implantable Pulse Generator, Model Number 3664
- Status
- Terminated
- Initiated
- 2018-03-09
- Posted
- 2018-03-23
- Terminated
- 2020-11-03
- Reason
- The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.
- Root cause
- Software change control
- Action
- Notification material in the form of a Customer Letter was hand-delivered to two consignees in the US beginning 09 Mar 2018. The Physician Letter was hand-delivered to impacted implanting physicians, beginning 14 Mar 2018. A Patient Letter was delivered via certified mailing to all active implanted patients in the US, beginning 14 Mar 2018.
- Quantity
- 17 units
- Distribution
- US
- Lot, serial or model codes
- Lot Number 6210849 Serial Number AVM174.1 Lot Number 6210852 Serial Numbers: AVA560.1, AVA214.1, AVL027.1 Lot Number 6499208 Serial Numbers: AVD426.1, AVA922.1 Lot Number 6263128 Serial Numbers: AVB119.1, AVD400.1, AVA622.1, AVD431.1, AVE633.1, AVK678.1, AVN837.1 Lot Number 6268723 Serial Numbers: AUV770.1, AVW805.1 Lot Number 6278155 Serial Number AVK896.1
- 510(k) numbers
- not on file
- PMA numbers
- P150004S002
- Product code
- PMP
- Firm FEI
- 1627487
- Firm city
- Plano
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28