FDA Medical Device Recalls (openFDA)

Exactech, Inc.: AcuMatch XLE Item Number / Product Description: 140-28-27 AM EHXL 0 LNR G 28MM 140-32-27 AM EHXL 0 LNR G 32MM 140-32-28 AM EHXL 0 LNR H 32MM 140-28-29 AM EHXL 0 LNR J 28MM 140-32-29 AM EHXL 0 LNR J 32MM 140-36-29 AM EHXL 0 LNR J 36MM 140-32-30 AM EHXL 0 LNR K 32MM 140-36-30 AM EHXL 0 LNR K 36MM 140-28-25 AM EHXL 0LNR E 28MM 140-28-26 AM EHXL 0LNR F 28MM 140-28-28 AM EHXL 0LNR H 28MM 140-36-28 AM EHXL 0LNR H 36MM 140-28-30 AM EHXL 0LNR K 28MM 142-28-24 AM EHXL 15 LNR D 28MM 142-28-26 AM EHXL 15 LNR F 28MM 142-32-27 AM EHXL 15 LNR G 32MM 142-32-28 AM EHXL 15 LNR H 32MM 142-36-28 AM EHXL 15 LNR H 36MM 142-36-29 AM EHXL 15 LNR J 26MM 142-28-29 AM EHXL 15 LNR J 28MM 142-32-29 AM EHXL 15 LNR J 32MM 142-28-30 AM EHXL 15 LNR K 28MM 142-32-30 AM EHXL 15 LNR K 32MM 142-36-30 AM EHXL 15 LNR K 36MM 142-28-25 AM EHXL 15LNR E 28MM 142-28-27 AM EHXL 15LNR G 28MM 142-28-28 AM EHXL 15LNR H 28MM 144-28-28 AM EHXL EXT LNR H 28MM 144-28-29 AM EHXL EXT LNR J 28MM 144-28-25 AM EHXL EXTLNR E 28MM 144-28-26 AM EHXL EXTLNR F 28MM 144-28-27 AM EHXL EXTLNR G 28MM 144-28-30 AM EHXL EXTLNR K 28MM 148-28-26 AM EHXL LAT LNR F 28MM 148-28-27 AM EHXL LAT LNR G 28MM 148-36-28 AM EHXL LAT LNR H 36MM 148-36-29 AM EHXL LAT LNR J 36MM 148-28-30 AM EHXL LAT LNR K 28MM 148-36-30 AM EHXL LAT LNR K 36MM 148-28-25 AM EHXL LATLNR E 28MM 148-28-28 AM EHXL LATLNR H 28MM 148-28-29 AM EHXL LATLNR J 28MM All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment.

Recall number
Z-1165-2025
Recalling firm
Exactech, Inc.
Product
AcuMatch XLE Item Number / Product Description: 140-28-27 AM EHXL 0 LNR G 28MM 140-32-27 AM EHXL 0 LNR G 32MM 140-32-28 AM EHXL 0 LNR H 32MM 140-28-29 AM EHXL 0 LNR J 28MM 140-32-29 AM EHXL 0 LNR J 32MM 140-36-29 AM EHXL 0 LNR J 36MM 140-32-30 AM EHXL 0 LNR K 32MM 140-36-30 AM EHXL 0 LNR K 36MM 140-28-25 AM EHXL 0LNR E 28MM 140-28-26 AM EHXL 0LNR F 28MM 140-28-28 AM EHXL 0LNR H 28MM 140-36-28 AM EHXL 0LNR H 36MM 140-28-30 AM EHXL 0LNR K 28MM 142-28-24 AM EHXL 15 LNR D 28MM 142-28-26 AM EHXL 15 LNR F 28MM 142-32-27 AM EHXL 15 LNR G 32MM 142-32-28 AM EHXL 15 LNR H 32MM 142-36-28 AM EHXL 15 LNR H 36MM 142-36-29 AM EHXL 15 LNR J 26MM 142-28-29 AM EHXL 15 LNR J 28MM 142-32-29 AM EHXL 15 LNR J 32MM 142-28-30 AM EHXL 15 LNR K 28MM 142-32-30 AM EHXL 15 LNR K 32MM 142-36-30 AM EHXL 15 LNR K 36MM 142-28-25 AM EHXL 15LNR E 28MM 142-28-27 AM EHXL 15LNR G 28MM 142-28-28 AM EHXL 15LNR H 28MM 144-28-28 AM EHXL EXT LNR H 28MM 144-28-29 AM EHXL EXT LNR J 28MM 144-28-25 AM EHXL EXTLNR E 28MM 144-28-26 AM EHXL EXTLNR F 28MM 144-28-27 AM EHXL EXTLNR G 28MM 144-28-30 AM EHXL EXTLNR K 28MM 148-28-26 AM EHXL LAT LNR F 28MM 148-28-27 AM EHXL LAT LNR G 28MM 148-36-28 AM EHXL LAT LNR H 36MM 148-36-29 AM EHXL LAT LNR J 36MM 148-28-30 AM EHXL LAT LNR K 28MM 148-36-30 AM EHXL LAT LNR K 36MM 148-28-25 AM EHXL LATLNR E 28MM 148-28-28 AM EHXL LATLNR H 28MM 148-28-29 AM EHXL LATLNR J 28MM All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment.
Status
Open, Classified
Initiated
2024-12-31
Posted
2025-02-12
Terminated
not on file
Reason
Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.
Root cause
Nonconforming Material/Component
Action
On 12/31/2024, the firm sent an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letter via FedEx to customers initially informing them that the firm was recalling specific hip polyethylene liner lots that were inadvertently packaged without the ethylene vinyl alcohol (EVOH) layer in the innermost bag. An updated customer notification letter was communicated to affected customer via email on 01/17/2025. Customers are instructed to: -Immediate Discontinuation: Immediately discontinue use of all affected products -Quarantine and Return: Segregate and quarantine affected products and arrange for their return to Exactech. The Recall Confirmation Form section of this communication provides product return instructions. Clinical Guidance: 1. Patient Monitoring: Surgeons should regularly monitor patients with affected devices for potential risks such as device wear, failure, component cracking or fracture, new or worsening pain, bone loss and/or swelling, as per the instructions for use. 2. Diagnostic Considerations: Diagnostic imaging, such as X-rays, may be considered if device failure is suspected. 3. Revision Considerations: Revision of well-functioning devices is not recommended for patients. 4. Resources: Healthcare providers and patients with questions can access educational resources on our website exac.com/recall-information and use the device serial look-up tool. Public Contact - 352-377-1140 or email matt.collins@exac.com
Quantity
42
Distribution
U.S.: AR, CA, CO, FL, GA, HI, IL, IN, KS, KY, LA, MD, MN, NV, NY, OH, OK, PR,SC, TN, TX and VA O.U.S.: Argentina, Colombia, Germany, Guatemala, Japan and Spain
Lot, serial or model codes
Item Number / Product Description/ UDI-DI code: 140-28-27 AM EHXL 0¿ LNR G 28MM 10885862537997 140-32-27 AM EHXL 0¿ LNR G 32MM 10885862538031 140-32-28 AM EHXL 0¿ LNR H 32MM 10885862538048 140-28-29 AM EHXL 0¿ LNR J 28MM 10885862538017 140-32-29 AM EHXL 0¿ LNR J 32MM 10885862538055 140-36-29 AM EHXL 0¿ LNR J 36MM 10885862538086 140-32-30 AM EHXL 0¿ LNR K 32MM 10885862538062 140-36-30 AM EHXL 0¿ LNR K 36MM 10885862538093 140-28-25 AM EHXL 0¿LNR E 28MM 10885862537973 140-28-26 AM EHXL 0¿LNR F 28MM 10885862537980 140-28-28 AM EHXL 0¿LNR H 28MM 10885862538000 140-36-28 AM EHXL 0¿LNR H 36MM 10885862538079 140-28-30 AM EHXL 0¿LNR K 28MM 10885862538024 142-28-24 AM EHXL 15¿ LNR D 28MM 10885862538123 142-28-26 AM EHXL 15¿ LNR F 28MM 10885862538147 142-32-27 AM EHXL 15¿ LNR G 32MM 10885862538192 142-32-28 AM EHXL 15¿ LNR H 32MM 10885862538208 142-36-28 AM EHXL 15¿ LNR H 36MM 10885862538239 142-36-29 AM EHXL 15¿ LNR J 26MM 10885862538246 142-28-29 AM EHXL 15¿ LNR J 28MM 10885862538178 142-32-29 AM EHXL 15¿ LNR J 32MM 10885862538215 142-28-30 AM EHXL 15¿ LNR K 28MM 10885862538185 142-32-30 AM EHXL 15¿ LNR K 32MM 10885862538222 142-36-30 AM EHXL 15¿ LNR K 36MM 10885862538253 142-28-25 AM EHXL 15¿LNR E 28MM 10885862538130 142-28-27 AM EHXL 15¿LNR G 28MM 10885862538154 142-28-28 AM EHXL 15¿LNR H 28MM 10885862538161 144-28-28 AM EHXL EXT LNR H 28MM 10885862538291 144-28-29 AM EHXL EXT LNR J 28MM 10885862538307 144-28-25 AM EHXL EXTLNR E 28MM 10885862538260 144-28-26 AM EHXL EXTLNR F 28MM 10885862538277 144-28-27 AM EHXL EXTLNR G 28MM 10885862538284 144-28-30 AM EHXL EXTLNR K 28MM 10885862538314 148-28-26 AM EHXL LAT LNR F 28MM 10885862538338 148-28-27 AM EHXL LAT LNR G 28MM 10885862538345 148-36-28 AM EHXL LAT LNR H 36MM 10885862538383 148-36-29 AM EHXL LAT LNR J 36MM 10885862538390 148-28-30 AM EHXL LAT LNR K 28MM 10885862538376 148-36-30 AM EHXL LAT LNR K 36MM 10885862538406 148-28-25 AM EHXL LATLNR E 28MM 10885862538321 148-28-28 AM EHXL LATLNR H 28MM 10885862538352 148-28-29 AM EHXL LATLNR J 28MM 10885862538369
510(k) numbers
["K173583"]
PMA numbers
not on file
Product code
OQI
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register