FDA Medical Device Recalls (openFDA)
CooperSurgical, Inc.: Milex Gellhorn Flexible under the following labels: 1) Gellhorn Flexible w/Short Stem Pessary, REF MXKPGSS2-1/4, 2-1/4" (57 mm) with Kit, 2) Gellhorn Flexible w/Short Stem Pessary, REF MXPGSS2-1/4, 2-1/4" (57 mm) without Kit, 3) Gellhorn Flexible w/Short Stem Pessary, REF MXPGSS2-3/4, 2-3/4" (70 mm)
- Recall number
- Z-1152-2021
- Recalling firm
- CooperSurgical, Inc.
- Product
- Milex Gellhorn Flexible under the following labels: 1) Gellhorn Flexible w/Short Stem Pessary, REF MXKPGSS2-1/4, 2-1/4" (57 mm) with Kit, 2) Gellhorn Flexible w/Short Stem Pessary, REF MXPGSS2-1/4, 2-1/4" (57 mm) without Kit, 3) Gellhorn Flexible w/Short Stem Pessary, REF MXPGSS2-3/4, 2-3/4" (70 mm)
- Status
- Terminated
- Initiated
- 2021-01-21
- Posted
- 2021-02-26
- Terminated
- 2024-01-03
- Reason
- The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print dimensions.
- Root cause
- Process change control
- Action
- On 01/21/2021, CooperSurgical issued Urgent Medical Device Recall notices to customers via certified mail. On 9/08/2021, CooperSurgical expanded the recall adding an additional lot and notified customers via courier service. Customers were advised to: - Discontinue use of and isolate all products subject to the recall. - Complete the Acknowledgement and Receipt Form, and return via email to recall@coopersurgical.com or fax to 203-601-9870. - Once the Acknowledgement Form is received, Cooper Surgical will arrange for Product replacement(s) and return of affected Product(s). The replacement Product(s) will be charged at the same price as the original order and your account will be credited back once the affected Product(s) are returned to CooperSurgical. If you do not have affected Product in inventory, please use the same Form to indicate that and return it to CooperSurgical. Customers with questions, please call 203-601-5200 ext. 3300.
- Quantity
- 109 units (Domestic: 60 units, Foreign: 49 units)
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, FL, IL, KS, MA, MI, NC, NY, PA, TN, TX, WA, & Puerto Rico. The countries of New Zealand, United Kingdom.
- Lot, serial or model codes
- Recall letter dated 1/21/2021: 1) With Kit: REF #: MXKPGSS2-1/4, Lot #: 276529- Exp Date June 10, 2025, 2) Without Kit: REF #: MXPGSS2-1/4, Lot #: 283083- Exp Date May 12, 2025; Lot 288491- Exp Date May 14, 2025; Lot # 288492- Exp Date May 29, 2025. Recall letter dated 9/8/2021 included one additional lot: 3) REF # MXPGSS2-3/4, Lot 276637- Exp Date February 12, 2026
- 510(k) numbers
- ["K904774"]
- PMA numbers
- not on file
- Product code
- HHW
- Firm FEI
- 1216677
- Firm city
- Trumbull
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28