Skip to the content

FDA Medical Device Recalls (openFDA)

Ethicon, Inc.: DERMABOND" PRINEO" Skin Closure System Product Usage: DERMABOND PRINEO System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND PRINEO System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound durinQ application of the liquid adhesive

Recall number
Z-1147-2018
Recalling firm
Ethicon, Inc.
Product
DERMABOND" PRINEO" Skin Closure System Product Usage: DERMABOND PRINEO System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND PRINEO System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound durinQ application of the liquid adhesive
Status
Terminated
Initiated
2017-11-27
Posted
not on file
Terminated
2019-04-30
Reason
Ethicon discovered that specific lots of DERMABOND"PRINEO"System may not dry within the specified time after proper application, and thus may fall off.
Root cause
Component change control
Action
Ethicon sent an URGENT MEDICAL DEVICE RECALL REMOVAL letter dated November 2017 to customers titled: "DERMABONDTM PRINEOTM SKIN CLOSURE SYSTEM (22 CM)". The letter identified the affected product, problem and action to be taken. The letter instructed customers to do not use or distribute any product which is subject to recall. Customers are asked to examine inventory, quarantine products that are subject to recall and contact the firm for return of product. Also, post the recall letter in a visible location at facility. If you have additional questions regarding this recall (removal) or to report any customer complaints, please contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 7:30 AM to 6:30 PM ET.
Quantity
20,090 eaches (10,545 units)
Distribution
US Nationwide Distribution
Lot, serial or model codes
LEJ230 LEJ246 LEJ259 LEJ368 LGP375 LGP605 LGP606 LGP675 LGP814 LGR689 LGR710 LGR756 LHH468 LHH469 LHH560 LHH608 LHH686 LHH784 LHP498 LHP599 LHP602 LHP868
510(k) numbers
["K133864"]
PMA numbers
not on file
Product code
OMO
Firm FEI
2210968
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register