Beckman Coulter Ireland, Inc.: The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.

Recall number
Z-1135-2026
Recalling firm
Beckman Coulter Ireland, Inc.
Product
The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.
Status
Open, Classified
Initiated
2025-12-08
Posted
2026-01-21
Terminated
not on file
Reason
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
Root cause
Device Design
Action
On 12/8/2025, recall notices were mailed and emailed to customers who were informed of the following: A clinically significant interference due to triglyceride concentration up to 1500mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL). Discontinuance or disposal of this product is not necessary. The LDL Cholesterol reagent can be used with the revised Instructions for Use (IFU) that contain the updated triglyceride interference specifications. Share the content of this letter with your laboratory and/or Medical Director. Retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. Complete and return the response form via email to regaffairsfax@beckman.com If you have any questions regarding this notice, please contact our Customer Support Center: http://www.beckmancoulter.com, or 1-800-854-3633 in the United States.
Quantity
34,451
Distribution
US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of
Lot, serial or model codes
REF/UDI-DI: OSR6196/15099590010928, OSR6296/15099590020361. All Lots and Expiration Dates. IFU Version BAOSR6X96-10 and prior.
510(k) numbers
K040471
PMA numbers
not on file
Product code
LBS
Firm FEI
1000206808
Firm city
Mills
Firm state
not on file
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register