Olympus America Inc.: Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4 --- Common name: Medical image storage device. Class I, 510 (k) exempt.

Recall number
Z-1126-03
Recalling firm
Olympus America Inc.
Product
Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4 --- Common name: Medical image storage device. Class I, 510 (k) exempt.
Status
Terminated
Initiated
2003-03-05
Posted
2003-08-12
Terminated
2005-03-22
Reason
Software defect; potential under certain circumstances to incorrectly identify endoscopy images that have been uploaded & saved to the network system.
Root cause
Other
Action
Olympus America, Inc. sent "Urgent-Device Correction" letters, dated 3/5/03, to the medical facilities, advising them of the software malfunction. The letter provides instructions to health care practitioners and that a new software release is being developed to prevent a recurrence.
Quantity
856
Distribution
Medical facilities nationwide and Puerto Rico; and 4 foreign accounts in Canada, Ecuador, Chile & Mexico.
Lot, serial or model codes
Versions 6.0 through 6.4
510(k) numbers
K934888, K934889
PMA numbers
not on file
Product code
LMB
Firm FEI
2429304
Firm city
Melville
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register