MEDLINE INDUSTRIES, LP - Northfield: Surgical convenience kits labeled as: a) CENTURION MEDICAL PRODUCTS CHEST TUBE INSERTION TRAY, Reorder Number CVI2105; b) CENTURION MEDICAL PRODUCTS VASECTOMY KIT, Reorder Number MNS11975; c) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT11670; d) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT12060; e) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT14705; f) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT17190; g) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT17195; h) CENTURION MEDICAL PRODUCTS LACERATION TRAY (P/S), Reorder Number SUT17755; i) CENTURION MEDICAL PRODUCTS LACERATION TRAY (P/S), Reorder Number SUT18475; j) CENTURION MEDICAL PRODUCTS SUTURE SET, Reorder Number SUT19505; k) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT19620; l) CENTURION MEDICAL PRODUCTS UNITYPOINT HEALTH LACERATION T, Reorder Number SUT20205; m) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT20410; n) CENTURION MEDICAL PRODUCTS BASIC SUTURE TRAY, Reorder Number SUT21360; o) CENTURION MEDICAL PRODUCTS SUTURE PACK TRAY, Reorder Number SUT4655; p) CENTURION MEDICAL PRODUCTS SUTURING SET, Reorder Number SUT5765; q) CENTURION MEDICAL PRODUCTS UMBILICAL CATH INSERT TRAY, NO, Reorder Number UVT1075A; r) CENTURION MEDICAL PRODUCTS UMBILICAL VESSEL CATH INSERT T, Reorder Number UVT505; s) CENTURION MEDICAL PRODUCTS UMBILICAL VESSEL TRAY, Reorder Number UVT835;

Recall number
Z-1119-2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Surgical convenience kits labeled as: a) CENTURION MEDICAL PRODUCTS CHEST TUBE INSERTION TRAY, Reorder Number CVI2105; b) CENTURION MEDICAL PRODUCTS VASECTOMY KIT, Reorder Number MNS11975; c) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT11670; d) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT12060; e) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT14705; f) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT17190; g) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT17195; h) CENTURION MEDICAL PRODUCTS LACERATION TRAY (P/S), Reorder Number SUT17755; i) CENTURION MEDICAL PRODUCTS LACERATION TRAY (P/S), Reorder Number SUT18475; j) CENTURION MEDICAL PRODUCTS SUTURE SET, Reorder Number SUT19505; k) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT19620; l) CENTURION MEDICAL PRODUCTS UNITYPOINT HEALTH LACERATION T, Reorder Number SUT20205; m) CENTURION MEDICAL PRODUCTS LACERATION TRAY, Reorder Number SUT20410; n) CENTURION MEDICAL PRODUCTS BASIC SUTURE TRAY, Reorder Number SUT21360; o) CENTURION MEDICAL PRODUCTS SUTURE PACK TRAY, Reorder Number SUT4655; p) CENTURION MEDICAL PRODUCTS SUTURING SET, Reorder Number SUT5765; q) CENTURION MEDICAL PRODUCTS UMBILICAL CATH INSERT TRAY, NO, Reorder Number UVT1075A; r) CENTURION MEDICAL PRODUCTS UMBILICAL VESSEL CATH INSERT T, Reorder Number UVT505; s) CENTURION MEDICAL PRODUCTS UMBILICAL VESSEL TRAY, Reorder Number UVT835;
Status
Open, Classified
Initiated
2022-12-07
Posted
2023-02-10
Terminated
not on file
Reason
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
Root cause
Under Investigation by firm
Action
The firm sent a correction notice to its direct consignees on 12/07/2022 by US mail and email. The notice explained the issue and requested the following: "Please check you stock for affected kits and over-labeling will be sent to place on kits to ensure that all labeling is accurate. No kits are to be returned back to Medline. All components are not affected. Our records indicate that these were used as a component in the list of kit(s) purchased from Medline Industries, LP." "If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP." "If you have any questions, please contact the Recall Department at 866-359-1704."
Quantity
2753 units
Distribution
Nationwide
Lot, serial or model codes
a) Reorder Number CVI2105, GTIN 40653160091546, Lot #22HMD248; b) Reorder Number MNS11975, GTIN 40653160325292, Lot #22HMD116; c) Reorder Number SUT11670, GTIN 40653160109762, Lot #22HMB913; d) Reorder Number SUT12060, GTIN 40653160120279, Lot #22HMC040; e) Reorder Number SUT14705, GTIN 40653160185438, Lot #22HMF267, 22HMF925; f) Reorder Number SUT17190, GTIN 40653160229231, Lot #22HMC041; g) Reorder Number SUT17195, GTIN 40653160229705, Lot #22HMH826; h) Reorder Number SUT17755, GTIN 40653160243701, Lot #22HMG491; i) Reorder Number SUT18475, GTIN 40653160258958, Lot #22HMH512; j) Reorder Number SUT19505, GTIN 40653160286166, Lot #22HMI064; k) Reorder Number SUT19620, GTIN 40653160289266, Lot #22HME946; l) Reorder Number SUT20205, GTIN 40653160293065, Lot #22IMA294; m) Reorder Number SUT20410, GTIN 40653160307052, Lot #22HME368; n) Reorder Number SUT21360, GTIN 40653160329436, Lot #22IMA124; o) Reorder Number SUT4655, GTIN 40653160015582, Lot #22HMC982; p) Reorder Number SUT5765, GTIN 40653160015865, Lot #22HMF927; q) Reorder Number UVT1075A, GTIN 10653160341956, Lot #22BBJ015; r) Reorder Number UVT505, GTIN 40653160182055, Lot #22FMF860; s) Reorder Number UVT835, GTIN 40653160289181, Lot #22GMF784
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRP
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register