FDA Medical Device Recalls (openFDA)

Beckman Coulter, Inc.: Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

Recall number
Z-1115-2025
Recalling firm
Beckman Coulter, Inc.
Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Status
Completed
Initiated
2024-01-03
Posted
2025-02-06
Terminated
not on file
Reason
A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.
Root cause
Software design
Action
A field service engineer was dispatched to the customer site to perform the software update and to replace the PCB TSI-RSI board from 01/03/2024 through 12/26/2024.
Quantity
61 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.
Lot, serial or model codes
UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.
510(k) numbers
K221225
PMA numbers
not on file
Product code
JJE
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register