Skip to the content

FDA Medical Device Recalls (openFDA)

Radiometer Medical ApS: Radiometer, REF:956-552, PIC050- Arterial Blood Sampler 1 X 2 mL, 80 IU electrolyte-balanced heparin. Not for injection. , IVD, Rx Only, Sterile EO, CE

Recall number
Z-1115-2022
Recalling firm
Radiometer Medical ApS
Product
Radiometer, REF:956-552, PIC050- Arterial Blood Sampler 1 X 2 mL, 80 IU electrolyte-balanced heparin. Not for injection. , IVD, Rx Only, Sterile EO, CE
Status
Open, Classified
Initiated
2022-04-08
Posted
not on file
Terminated
not on file
Reason
Arterial Blood Sampler sterility cannot be guaranteed throughout the shelf life of the product.
Root cause
Package design/selection
Action
On April 26, 2022 Radiometer issued a "Urgent Medical Device Recall" notification via Email. The firm ask consignees to take the following actions: Radiometer kindly request you to stop using the affected product with immediate effect. To ensure patient safety in your facility, kindly follow the steps below: " Check your inventory of the above PICO50 Arterial Blood Samplers " Check for the above PICO50 Arterial Blood Samplers distributed in your institution. " Collect any of the above PICO50 Arterial Blood Samplers and put on quarantine " Complete the Recall Response Form (last page of this letter) and return to your Radiometer representative together with the quarantined PICO50 Arterial Blood Samplers. To ensure you receive credit for the PICO50 Arterial Blood Samplers, please contact Customer Service @ CSRsupport@radiometeramerica.com to receive a Return Goods Authorization
Quantity
7,228 boxes (100 units per box)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DC, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NM, NY, OH, OK, OR, PA, TN, TX, WA, WI and the countries of Canada, Italy, Norway, Austria, Algeria, Argentina, Bolivia, Bosnia, Brazil, Burkina, Faso, Chile, Colombia, Cyprus, Ecuador, Egypt, Eswatin, Ethiopia, Greece, Guatemala, Israel, Lebanon, Libya, Lithuania, Malawi, Nicaragua, Romania, Senegal, Serbia, Slovenia, Thailand, Uganda, Uruguay, Vietnam, Zimbabwe, Belgium, Netherlands, China, Macao, Croatia, Hungary, Denmark, Iceland, Germany, Spain, Portugal, India, Japan, United Kingdom, Ireland, Russia, Australia, Switzerland, France, Czech Republic, Slovakia, South Africa, Botswana, Namibia, Poland, Turkey, Georgia, Sweden, Finland, Estonia, Kazakhstan, Malaysia, Hong Kong.
Lot, serial or model codes
"All Lots" UDI/DI: 05700699565525
510(k) numbers
["K962158"]
PMA numbers
not on file
Product code
JKA
Firm FEI
3002807968
Firm city
Bronshoj
Firm state
not on file
Firm country
Denmark

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register