Skip to the content

FDA Medical Device Recalls (openFDA)

Edwards Lifesciences, LLC: Edwards Lifesciences(R) IntraClude(TM) Intra-Aortic Occlusion Device is a 10.5 Fr (3.5mm), triple lumen, 100cm long catheter with an elastomeric balloon near its distal tip designed to occlude the ascending aorta in order to partition the aortic root from arterial circulation. The balloon expands to occlude a range of aorta sizes from 20 to 40 mm. The large central lumen of the IntraClude Device serves three functions: accommodating the guidewire, delivering cardioplegia solution to the aortic root, and venting fluid and air from the aortic root. The two remaining lumens serve as conduits for balloon inflation/deflation and aortic root pressure monitoring.

Recall number
Z-1100-2019
Recalling firm
Edwards Lifesciences, LLC
Product
Edwards Lifesciences(R) IntraClude(TM) Intra-Aortic Occlusion Device is a 10.5 Fr (3.5mm), triple lumen, 100cm long catheter with an elastomeric balloon near its distal tip designed to occlude the ascending aorta in order to partition the aortic root from arterial circulation. The balloon expands to occlude a range of aorta sizes from 20 to 40 mm. The large central lumen of the IntraClude Device serves three functions: accommodating the guidewire, delivering cardioplegia solution to the aortic root, and venting fluid and air from the aortic root. The two remaining lumens serve as conduits for balloon inflation/deflation and aortic root pressure monitoring.
Status
Open, Classified
Initiated
2019-01-16
Posted
not on file
Terminated
not on file
Reason
Edwards Lifesciences has received a limited number of customer reports regarding an inter-lumen leak between the cardioplegia lumen and pressure lumen in the device.
Root cause
Process control
Action
The firm Edwards Lifesciences, disseminated the notices on 01/16/2019 by letter entitled "URGENT PRODUCT NOTIFICATION ACTION REQUIRED". The letter stated the problem, how to identify it, and actions to be taken. The customers were instructed to do the following: 1. Review this field safety notice to understand the potential hazard. 2. Meet and review with appropriate clinical staff at your institution. 3. Complete and return the acknowledgement form attached to this letter via fax to Fax: 949.250.2525 or email to www.edwards.com within five (5) business days of receiving this notice. 4. Distribute this notice within your organization or to any organization where the potentially affected devices may have been transferred. 5. Product may continue to be used. Product return is not required. If you have questions that have not been answered by this letter, please call Edwards Customer Service Monday through Friday at 949.250.2500.
Quantity
2715 units
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Czech Republic, France Germany, Israel, Italy, Luxembourg, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, UAE, and United Kingdom.
Lot, serial or model codes
60850886 60850888 60850885 60850889 60850890 60876468 60760965 60887217 60887219 60887220 60934623 60887218 60934624 60930494 60934625 60972889 60972890 60972892 61014406 61014408 61014410 61014411 61050171 61050172 61078032 61078031 61097633 61139240 61139239 61139241 61139242 61184965 61184964 61184966 61184967 61204586 61204587 61259626 61259627 61259628 61259629 61288501 61369214 61369215 61369216 61369217 61425732 61425734 61425733 61495692 61517743 61495693 61531653 61531655 61704542 61531654 61704656 61713205 61713206 61713208 61713211 61713207 61713209 61713210 61713216 61713217 61713212 61713214 61713213 61713215
510(k) numbers
["K163693"]
PMA numbers
not on file
Product code
DXC
Firm FEI
2015691
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register