Skip to the content

FDA Medical Device Recalls (openFDA)

Diagnostica Stago, Inc.: Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.

Recall number
Z-1084-2026
Recalling firm
Diagnostica Stago, Inc.
Product
Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.
Status
Open, Classified
Initiated
2025-12-15
Posted
2026-01-14
Terminated
not on file
Reason
Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
Root cause
Under Investigation by firm
Action
Diagnostica Stago notified consignees of the recall on about 12/15/2025 via email. Consignees were instructed to stop using Lot 271288 and dispose of all remaining product according to local regulations, and complete, sign, and return the attached Acknowledgement Form provided. Additionally, consignees were instructed that if they have distributed the affected lot to any other laboratory, please notify those recipients of this recall and ask them to complete and return an Acknowledgement Form.
Quantity
438 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, HI, IA, MA, MI, NJ, NV, NY, OH, TN, WI and the countries of Belgium, Canada, Denmark, Algeria, Spain, Finland, France, Great Britain, Italy, Lebanon, Saudi Arabia, Uganda.
Lot, serial or model codes
Reference Number: 00615. UDI: (01)03607450006155(11)240531(17)260531(10)271288(241)00615. Lot Number: 271288.
510(k) numbers
["K003767"]
PMA numbers
not on file
Product code
LCO
Firm FEI
2245451
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06