Elekta, Inc.: RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Recall number
Z-1081-2014
Recalling firm
Elekta, Inc.
Product
RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Status
Terminated
Initiated
2013-11-25
Posted
2014-03-06
Terminated
2015-08-13
Reason
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Root cause
Software design
Action
Elekta sent an Important Field Safety Notice November 2013 to all affected customers. Corrective Action #1 - Release of Important Field Safety Notice 200 01 502 041 - "Recommended quality assurance after a calibration procedure" Released 25 November, 2013. Important Field Safety Notice is in the process of being distributed to customers advising them of the recommended Quality Assurance checks to avoid/detect the problem. The permanent solution to this problem is currently been developed. Corrective Action #2 - Permanent Solution - All users of Desktop Pro software will be informed of a mandatory upgrade to Integrity 1.2 upon its release which will prevent calibration from being inverted. For questions regarding this recall call 770-300-9725.
Quantity
489
Distribution
Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.
Lot, serial or model codes
R4.2, R5.0, R5.1, R6.0, R6.1
510(k) numbers
K982713
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register