Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) IV START KIT NS, Medline kit SKU DYKM1980C; 2) (134) L&D IV-START SET, Medline kit SKU DYKM2137B; 3) IV START KIT, Medline kit SKU DYKS1372C; 4) IV START KIT, Medline kit SKU DYNDV1691B; 5) OUTPATIENT 2 TO 5 YEARS, Medline kit SKU DYNDV2464A.

Recall number
Z-1056-2026
Recalling firm
Medline Industries, LP
Product
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) IV START KIT NS, Medline kit SKU DYKM1980C; 2) (134) L&D IV-START SET, Medline kit SKU DYKM2137B; 3) IV START KIT, Medline kit SKU DYKS1372C; 4) IV START KIT, Medline kit SKU DYNDV1691B; 5) OUTPATIENT 2 TO 5 YEARS, Medline kit SKU DYNDV2464A.
Status
Open, Classified
Initiated
2025-11-26
Posted
2026-01-12
Terminated
not on file
Reason
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Root cause
Nonconforming Material/Component
Action
Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
5314 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Medline kit SKU DYKM1980C, UDI/DI 10195327608309 (EA) 40195327608300 (CS), Lot Number 25JBB147; Medline kit SKU DYKM1980C, UDI/DI 10195327608309 (EA) 40195327608300 (CS), Lot Number 25KBL305; Medline kit SKU DYKM2137B, UDI/DI 10198459285615 (EA) 40198459285616 (CS), Lot Number 25HBT834; Medline kit SKU DYKM2137B, UDI/DI 10198459285615 (EA) 40198459285616 (CS), Lot Number 25IBJ757; Medline kit SKU DYKM2137B, UDI/DI 10198459285615 (EA) 40198459285616 (CS), Lot Number 25JBJ557; Medline kit SKU DYKS1372C, UDI/DI 10195327373504 (EA) 40195327373505 (CS), Lot Number 25JLA480; Medline kit SKU DYNDV1691B, UDI/DI 10195327658472 (EA) 40195327658473 (CS), Lot Number 25HBI335; Medline kit SKU DYNDV1691B, UDI/DI 10195327658472 (EA) 40195327658473 (CS), Lot Number 25IBE523; Medline kit SKU DYNDV1691B, UDI/DI 10195327658472 (EA) 40195327658473 (CS), Lot Number 25JBM702; Medline kit SKU DYNDV2464A, UDI/DI 10195327065911 (EA) 40195327065912 (CS), Lot Number 25IBV942.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRS
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register