FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: Stryker Impaction Bur Guard REF 2296-301, 2296-300, 0260-901-300, & 5400-300; Stryker Instruments, Kalamazoo, MI. Intended to be used with impaction Drill. Drill is used for surgical procedures involving precision drilling and cutting of bone and hard tissue. The bur guard covers the nose of the handpiece and the shaft of the cutting accessory to prevent exposure to heat and moving parts, as well as minimize bur whip and handpiece vibration.
- Recall number
- Z-1052-2009
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- Stryker Impaction Bur Guard REF 2296-301, 2296-300, 0260-901-300, & 5400-300; Stryker Instruments, Kalamazoo, MI. Intended to be used with impaction Drill. Drill is used for surgical procedures involving precision drilling and cutting of bone and hard tissue. The bur guard covers the nose of the handpiece and the shaft of the cutting accessory to prevent exposure to heat and moving parts, as well as minimize bur whip and handpiece vibration.
- Status
- Terminated
- Initiated
- 2009-01-27
- Posted
- 2009-03-30
- Terminated
- 2009-12-17
- Reason
- Bur guards were distributed without an expiration date, and if they are used after they have expired it could result in the patient or operating room staff experiencing a burn or laceration.
- Root cause
- Process control
- Action
- Customers were notified via a Medical Device Recall Notification letter dated 1/27/09 and instructed to immediately check all stock areas or operating room storage and inspect their bur guards for an expiration date. They were also instructed to quarantine any affected product found, to indicate on the Business Reply Form the quantity of affected product on hand, to destroy any affected product found without an expiration date by breaking off at least 3 of the tabs on the bur guard, then dispose of the affected product in normal trash, and complete and return the business reply form. Customers are to fax the Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2760. Upon receipt of the Business Reply Form, a replacement bur guard(s) will be sent for any affected product destroyed.
- Quantity
- 70
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Any hand pieces repaired and returned from 9/18/08 to 10/10/08 may have the recalled bur guards. Serial Numbers: 0414100533, 0414100563, 0509100113, 0509100153, 0723404443, 0719104803, 98030843, 0318200043, 0727604043, 0729800243, 98061133, 0329401103, 97060133, 0424700493, 95022553, 0225501513, 0326900083, 99020483, 0330800593, 96010273, 0318200013, 97010813, 0324102133, 96050103, 0225501563, 99050023, 94091003, 97060223, 97060243, 97060613, 97080253, 98080163, 98050533, 98020403, 0533401073, 97050523, 0428800503, 0713801143, 0633303273, 96070353, 0311401713, 00120423, 99050053, 01020103, 02010803, 96120423, 99090353, 0233004783, 98061203, 99090293, 99110863, 0428202963, 0624803663, 0719000333, 94121523, 99100013, 94080453, 97050833, 96090533, 97120983, 97120993, 99111173, 0726300403, 0726300413, 01070163, 01070173, 0409604363, 0418005013, 0528001463, 99010543, 0511800053, 0511800083, 0709208133, 98050813, 00020373, 97020173, 99100123, 97050823, 01040533, 0228902373, 0231700673, 99070473, 0313302433, 99010903, 0819605873, 95040843, 94120653, 0715001683, 95021513, 95021523, 98010663, 0230100243, 98100133, 99100073, 0713123243, 0415500963, 95011113, 0726708833, 95030323, 0230100233, 99090493, 00080073, 95100103, 95100113, 0230200313, 97040563, 99070123, 0316200853, 94090983, 00110113, 94111083, 99010683, 97110023, 98120993, 00030153, 0512301433, 94100163, 94121103, 94121113, 98010673, 96110383, 97110453, 0330700333, 97121623, 98100863, 0414900263, 0232500443, 95022133, 0414100543, 0414100553, 0414100613, 0509100133, 0723404483, 98111233, 95100203, 0734700603, 00080083, 98050053, 00120433, 99040403, 99100113, 0309902613, 0808801833, 0233801463, 0721406933, 94121903, 99080283, 0731614433, 0608304673, 0609004613, 0611506303, 0611506313, 0613003513, 0613003533, 0525500823, 96120233, 95110333, 0417601883, 0618101713, 0624205473, 94090153, 98120473, 97100113, 0233900963, 00100793, 00030163, 0225600983, 0231000193, 96020173, 95030143, 01090313, 0606604453, 0334600733, 98050113, 0231800503, 94120733, 94050013, 0715001663, 94122153, 95030993, 95120103, 94110683, 0635406433, 98081583, 99020373, 94070313, 99111033, 0315500203, 96070313, 0535500863, 0632200223, 0417601873, 0400500163, 0631003513, 97080383, 0621206363, 02010793, 0401402363, 0336400963, 0330901153, 0229100103, 0508201013, 0634605013, 0635406413, 95100603, 00060133, 0530602873, 98011083, 98021223, 97111273, 98120113, 97030263, 97050393, 97020163, 99100253, 97090053, 00040293, 0703902843, 0707903723, 97050563, and 97090943.
- 510(k) numbers
- K961970
- PMA numbers
- not on file
- Product code
- HWE
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- 3e6df46828c1
- Size
- 288213574 bytes
- Published
- 2026-10-06
- Read
- 2026-10-07T03:20:25.099Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-14
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06