FDA Medical Device Recalls (openFDA)

Carl Zeiss Meditec AG: FORUM Archive and Viewer, version 3.1, v 3.1.1, (DVD Format) and v 3.2, v 3.2.1.(DVD and USB Flash Drive Format). Catalog numbers: 000000-20107-750 (DVD with either FORUM v 3.1 or v 3.1.1) 000000-2058-601 (DVD with either FORUM v 3.2 or 3.2.1); 000000-2084-928 (USB drive with FORUM 3.2.1) Ophthalmic image management system.

Recall number
Z-1049-2015
Recalling firm
Carl Zeiss Meditec AG
Product
FORUM Archive and Viewer, version 3.1, v 3.1.1, (DVD Format) and v 3.2, v 3.2.1.(DVD and USB Flash Drive Format). Catalog numbers: 000000-20107-750 (DVD with either FORUM v 3.1 or v 3.1.1) 000000-2058-601 (DVD with either FORUM v 3.2 or 3.2.1); 000000-2084-928 (USB drive with FORUM 3.2.1) Ophthalmic image management system.
Status
Terminated
Initiated
2015-01-22
Posted
2015-02-04
Terminated
2016-03-04
Reason
Software defect in the FORUM Viewer versions 3.1 and 3.2 which may lead to misinterpretation of the optical coherence tomography (OCT) data.
Root cause
Software in the Use Environment
Action
Important Field Corrective Action for Forum Viewer letters were sent to all affected US customers on January 22, 2015 by certified mail.
Quantity
985
Distribution
Nationwide Distribution.
Lot, serial or model codes
Catalog numbers: 000000-20107-750 (DVD with either FORUM v 3.1 or v 3.1.1) 000000-2058-601 (DVD with either FORUM v 3.2 or 3.2.1); 000000-2084-928 (USB drive with FORUM 3.2.1)
510(k) numbers
["K122938"]
PMA numbers
not on file
Product code
NFJ
Firm FEI
3001362763
Firm city
Jena
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register