FDA Medical Device Recalls (openFDA)
Immuno Concepts Inc: ANA Fluorescent Test System and ANA Colorzyme Test System, Sub assembly catalog number SA 2013, Kit Catalog Numbers, 2000, 2002, 2003, 2004, 2005, 2100, 2200, 4000, 4200, 8024.1, and 8060, manufactured by Immuno Concepts N.A. Ltd, Sacramento, CA. The device is used in the screening and titration of circulating antinuclear antibodies by means of an immunofluoresence test. Products are to be used as an aid in the detection of antibodies associated with systemic rheumatic disease.
- Recall number
- Z-1044-2009
- Recalling firm
- Immuno Concepts Inc
- Product
- ANA Fluorescent Test System and ANA Colorzyme Test System, Sub assembly catalog number SA 2013, Kit Catalog Numbers, 2000, 2002, 2003, 2004, 2005, 2100, 2200, 4000, 4200, 8024.1, and 8060, manufactured by Immuno Concepts N.A. Ltd, Sacramento, CA. The device is used in the screening and titration of circulating antinuclear antibodies by means of an immunofluoresence test. Products are to be used as an aid in the detection of antibodies associated with systemic rheumatic disease.
- Status
- Terminated
- Initiated
- 2008-12-04
- Posted
- 2009-04-06
- Terminated
- 2009-11-10
- Reason
- Samples can occasionally run out of the wells and cross contaminate adjacent wells.
- Root cause
- Nonconforming Material/Component
- Action
- The firm notified customers via a "Notice of Market Withdrawal" letter, e-mail and telephone. The letter described the problem and instructed customers to removed affected product from their inventory, complete the attached form and return it via fax to Immuno Concepts, Inc. at 916-363-2843. The firm will issue replacements as requested. Direct questions to Immuno Concepts, Inc. at 916-363-2649.
- Quantity
- 378 kits
- Distribution
- Worldwide Distribution, USA (NC, WI, MS, MD, IL, GA, CA, MI and Puerto Rico) and countries of Sweden, Mexico, Spain and Denmark.
- Lot, serial or model codes
- Sub Assembly Lot Numbers: 0823200, 0823201, and 0825500 and Kit Lot Numbers: 826305, 0827311, 0828107, 0828108, 0828109, 0830914, 0831607, 0831606, 0829719, 0827315, 0827301, 0829413, 0829512, 0826905, 0828114, 0828804, 0829414, 0829721, 0830300, 0828307 and 0829513.
- 510(k) numbers
- ["K803276","K833148"]
- PMA numbers
- not on file
- Product code
- DHN
- Firm FEI
- 2918768
- Firm city
- Sacramento
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as