Skip to the content

FDA Medical Device Recalls (openFDA)

Immuno Concepts Inc: ANA Fluorescent Test System and ANA Colorzyme Test System, Sub assembly catalog number SA 2013, Kit Catalog Numbers, 2000, 2002, 2003, 2004, 2005, 2100, 2200, 4000, 4200, 8024.1, and 8060, manufactured by Immuno Concepts N.A. Ltd, Sacramento, CA. The device is used in the screening and titration of circulating antinuclear antibodies by means of an immunofluoresence test. Products are to be used as an aid in the detection of antibodies associated with systemic rheumatic disease.

Recall number
Z-1044-2009
Recalling firm
Immuno Concepts Inc
Product
ANA Fluorescent Test System and ANA Colorzyme Test System, Sub assembly catalog number SA 2013, Kit Catalog Numbers, 2000, 2002, 2003, 2004, 2005, 2100, 2200, 4000, 4200, 8024.1, and 8060, manufactured by Immuno Concepts N.A. Ltd, Sacramento, CA. The device is used in the screening and titration of circulating antinuclear antibodies by means of an immunofluoresence test. Products are to be used as an aid in the detection of antibodies associated with systemic rheumatic disease.
Status
Terminated
Initiated
2008-12-04
Posted
2009-04-06
Terminated
2009-11-10
Reason
Samples can occasionally run out of the wells and cross contaminate adjacent wells.
Root cause
Nonconforming Material/Component
Action
The firm notified customers via a "Notice of Market Withdrawal" letter, e-mail and telephone. The letter described the problem and instructed customers to removed affected product from their inventory, complete the attached form and return it via fax to Immuno Concepts, Inc. at 916-363-2843. The firm will issue replacements as requested. Direct questions to Immuno Concepts, Inc. at 916-363-2649.
Quantity
378 kits
Distribution
Worldwide Distribution, USA (NC, WI, MS, MD, IL, GA, CA, MI and Puerto Rico) and countries of Sweden, Mexico, Spain and Denmark.
Lot, serial or model codes
Sub Assembly Lot Numbers: 0823200, 0823201, and 0825500 and Kit Lot Numbers: 826305, 0827311, 0828107, 0828108, 0828109, 0830914, 0831607, 0831606, 0829719, 0827315, 0827301, 0829413, 0829512, 0826905, 0828114, 0828804, 0829414, 0829721, 0830300, 0828307 and 0829513.
510(k) numbers
["K803276","K833148"]
PMA numbers
not on file
Product code
DHN
Firm FEI
2918768
Firm city
Sacramento
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register