Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD BBL Sensi Disc Tetracycline 30 ug,used for semi-quantitative in vitro susceptibility testing. Catalog Number 230998 and 231344

Recall number
Z-1040-2024
Recalling firm
Becton Dickinson & Co.
Product
BD BBL Sensi Disc Tetracycline 30 ug,used for semi-quantitative in vitro susceptibility testing. Catalog Number 230998 and 231344
Status
Open, Classified
Initiated
2024-01-08
Posted
2024-02-08
Terminated
not on file
Reason
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Root cause
Process control
Action
On January 8, 2024, the firm notified customers through URGENT Medical Device Product Correction letters. Users were instructed to cease use of catalog number 291270, and to refrain from conducting AST for H. influenzae with impacted Sensi-Discs. Customers should destroy any units of catalog number 291270 in their inventory. Customers should share and post the recall notification within their facility network and forward to any customers to whom they may have distributed the product. Future lots of affected Sensi-Disc products (excluding catalog number 291270) will be labeled with a sticker notifying customers "This product should not be used for the semi-quantitative in vitro susceptibility testing of Haemophilus influenzae." Identify impacted lot numbers: https://legacy.bd.com/alerts-notices/IDS-23-4851.asp For customer questions: Phone: 1-844-823-5433 Say "Recall when prompted Mon-Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Quantity
2,363,168 total units
Distribution
Domestic distribution nationwide. International distribution worldwide.
Lot, serial or model codes
Catalog No. 230998 UDI-DI 00382902309987 Lots 1242880 1277918 1301570 2025088 2056233 2118569 2276681 3094187 3212760 3263267 9009914 9046895 0065366 0178210; Catalog No. 231344¿ UDI-DI 30382902313442 Lots 3010027 3062263 3094187 3158029 3212760 3263267 3275644 0065366 0085202 0230453 0274792 0335906 1095603 1148539 1180271 1242880 1277918 1301570 2025088 2056233 2089874 2118569 2276681 8355562 9009914 9046895 9130797 9239639 9263061
510(k) numbers
not on file
PMA numbers
not on file
Product code
JTN
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register