Guidant Corporation: CONTAK TR (Model no. 1241) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. The Guidant CONTAK TR CHFD congestive heart failure devices are multiprogrammable dual-chamber devices that offer adaptive-rate congestive heart failure pacing therapy. CONTAK TR devices use an accelerometer that allows the pacing rate to adapt to the patient's changing metabolic demand as it responds to patient activity.

Recall number
Z-1039-05
Recalling firm
Guidant Corporation
Product
CONTAK TR (Model no. 1241) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. The Guidant CONTAK TR CHFD congestive heart failure devices are multiprogrammable dual-chamber devices that offer adaptive-rate congestive heart failure pacing therapy. CONTAK TR devices use an accelerometer that allows the pacing rate to adapt to the patient's changing metabolic demand as it responds to patient activity.
Status
Terminated
Initiated
2005-07-18
Posted
2005-07-28
Terminated
2007-01-05
Reason
A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
Root cause
Other
Action
INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.
Quantity
INITIAL RECALL 164. EXPANDED RECALL 110.
Distribution
worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.
Lot, serial or model codes
Affected devices may be contained in the following serial number ranges.  INITIAL RECALL serial numbers 200128 thru 200479. EXPANDED RECALL serial numbers 200076 thru 200958
510(k) numbers
not on file
PMA numbers
["P940031"]
Product code
LWP
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register