Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: OSS Modular Arthrodesis Nail, 7 Degree Collar Assembly with Locking Bolts-For use in the replacement of segmental defects of long bone, including midshaft replacement, and arthrodesis of the knee Item Number: CP260602

Recall number
Z-1036-2023
Recalling firm
Biomet, Inc.
Product
OSS Modular Arthrodesis Nail, 7 Degree Collar Assembly with Locking Bolts-For use in the replacement of segmental defects of long bone, including midshaft replacement, and arthrodesis of the knee Item Number: CP260602
Status
Open, Classified
Initiated
2022-11-29
Posted
2023-01-27
Terminated
not on file
Reason
Product missing the four small locking collar bolts resulting in extension of surgery to find a replacement
Root cause
Process control
Action
****UPDATE***Phase II Zimmer Biomet issued Urgent Medical Device Recall Letter (Removal) on March 7, via FedEx to the Distributor and Risk Manager. Letter states health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Zimmer Biomet is conducting the recall in a phased approach due to limited replacement products for the Modular Arthrodesis Nail Collar Assembly (PN: CP260602) and due to the ability to separately purchase the small locking collar bolts (PN: CP260603).Letter to Distributors via email on 11/29/22 states: Replacement Small Locking Collar Bolts (Part Number: CP260603) will be sent to your territory. Replacement orders were placed today, November 29, with overnight delivery to your main territory office (Phase 1) Once replacement devices are availab
Quantity
2 units
Distribution
US Nationwide distribution in the states of CA, KY, FL.
Lot, serial or model codes
UDI: (01)00880304082991(17)321103(10)136620 Lot Number: 136620
510(k) numbers
["K042409"]
PMA numbers
not on file
Product code
HSB
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register