Skip to the content

FDA Medical Device Recalls (openFDA)

Avanos Medical, Inc.: Avanos Cortrak 2 Enteral Access System (EAS)

Recall number
Z-1032-2022
Recalling firm
Avanos Medical, Inc.
Product
Avanos Cortrak 2 Enteral Access System (EAS)
Status
Terminated
Initiated
2022-03-21
Posted
2022-05-13
Terminated
2024-09-09
Reason
Product labeling is being modified. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol.
Root cause
Device Design
Action
Avanos Medical issued URGENT: FIELD CORRECTION Notices dated March 18 and March 21, 2022. Phase 2 of the field action will occur later. The Notices instructed consignees to: CHECK all storage and usage locations to determine if any impacted product remains within your possession. ATTACH THIS LETTER to the Operators Manual until a revised Operators Manual, Quick Start Guide and Troubleshooting Tip Cards are available. COMPLETE and RETURN the Acknowledgement Form to Avanos, via email to avanos-fca002@iqvia.com, Phone Number (855) 365-3981. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol. USE extreme caution in patients who are combative or who move excessively during placement as the receiver unit may move, impacting the placement tracing. If a deviation from the midline is seen on the All-In-One Monitor display during advancement in the upper quadrants, or resistance is encountered: 1) STOP advancement of the system. 2) REMOVE the tube and stylet and assess the patient for injury per institutional protocol Device should only be used by qualified, trained users. If you need additional training, please contact your local field sales representative. Phase 2 letters were sent via FedEx to all consignees on 09/27/2022 and included physical copies of the updated versions of the Operator's Manual, Quick Reference Guide, and Troubleshooting Tips.
Quantity
721 devices
Distribution
Worldwide Distribution. US Nationwide, Australia, Austria, Canada, Germany, Ireland, Kuwait, New Zealand, India, Japan, Malaysia, Netherlands, Panama, Singapore, Switzerland, and United Kingdom.
Lot, serial or model codes
1) Product Code 20-0950, CORTRAK 2 Enteral Access System (EAS), UDI: 00350770472010; all serial numbers 2) Product Code P20-0950, CORTRAK 2 Enteral Access System (EAS) - Loaner Unit, UDI: 00350770472065; all serial numbers 3) Product Code 20-0950, CORTRAK 2 Enteral Access System (EAS) - Halyard version; UDI: 10680651472011; all serial numbers 4) Product Code P20-0950, CORTRAK 2 Enteral Access System (EAS) - Loaner Unit - Halyard version; UDI: 10680651472066; all serial numbers
510(k) numbers
["K113351","K191340"]
PMA numbers
not on file
Product code
KNT
Firm FEI
3011270181
Firm city
Alpharetta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28