Avanos Medical, Inc.: Avanos Cortrak 2 Enteral Access System (EAS)
- Recall number
- Z-1032-2022
- Recalling firm
- Avanos Medical, Inc.
- Product
- Avanos Cortrak 2 Enteral Access System (EAS)
- Status
- Terminated
- Initiated
- 2022-03-21
- Posted
- 2022-05-13
- Terminated
- 2024-09-09
- Reason
- Product labeling is being modified. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol.
- Root cause
- Device Design
- Action
- Avanos Medical issued URGENT: FIELD CORRECTION Notices dated March 18 and March 21, 2022. Phase 2 of the field action will occur later. The Notices instructed consignees to: CHECK all storage and usage locations to determine if any impacted product remains within your possession. ATTACH THIS LETTER to the Operators Manual until a revised Operators Manual, Quick Start Guide and Troubleshooting Tip Cards are available. COMPLETE and RETURN the Acknowledgement Form to Avanos, via email to avanos-fca002@iqvia.com, Phone Number (855) 365-3981. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol. USE extreme caution in patients who are combative or who move excessively during placement as the receiver unit may move, impacting the placement tracing. If a deviation from the midline is seen on the All-In-One Monitor display during advancement in the upper quadrants, or resistance is encountered: 1) STOP advancement of the system. 2) REMOVE the tube and stylet and assess the patient for injury per institutional protocol Device should only be used by qualified, trained users. If you need additional training, please contact your local field sales representative. Phase 2 letters were sent via FedEx to all consignees on 09/27/2022 and included physical copies of the updated versions of the Operator's Manual, Quick Reference Guide, and Troubleshooting Tips.
- Quantity
- 721 devices
- Distribution
- Worldwide Distribution. US Nationwide, Australia, Austria, Canada, Germany, Ireland, Kuwait, New Zealand, India, Japan, Malaysia, Netherlands, Panama, Singapore, Switzerland, and United Kingdom.
- Lot, serial or model codes
- 1) Product Code 20-0950, CORTRAK 2 Enteral Access System (EAS), UDI: 00350770472010; all serial numbers 2) Product Code P20-0950, CORTRAK 2 Enteral Access System (EAS) - Loaner Unit, UDI: 00350770472065; all serial numbers 3) Product Code 20-0950, CORTRAK 2 Enteral Access System (EAS) - Halyard version; UDI: 10680651472011; all serial numbers 4) Product Code P20-0950, CORTRAK 2 Enteral Access System (EAS) - Loaner Unit - Halyard version; UDI: 10680651472066; all serial numbers
- 510(k) numbers
- ["K113351","K191340"]
- PMA numbers
- not on file
- Product code
- KNT
- Firm FEI
- 3011270181
- Firm city
- Alpharetta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28