Guidant Corporation: DISCOVERY (model nos. 1174, 1175, 1273, 1274 and 1275) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. DISCOVERY pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy. DISCOVERY pacemakers use an accelerometer that allows the pacing rate to adapt to the patient's changing metabolic demand as it responds to patient activity.

Recall number
Z-1032-05
Recalling firm
Guidant Corporation
Product
DISCOVERY (model nos. 1174, 1175, 1273, 1274 and 1275) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. DISCOVERY pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy. DISCOVERY pacemakers use an accelerometer that allows the pacing rate to adapt to the patient's changing metabolic demand as it responds to patient activity.
Status
Terminated
Initiated
2005-07-18
Posted
2005-07-28
Terminated
2007-01-05
Reason
A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
Root cause
Other
Action
INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.
Quantity
INITIAL RECALL 14,265. EXPANDED RECALL 12,876
Distribution
worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.
Lot, serial or model codes
Affected devices may be contained in the following serial number ranges.  Model 1174: INITIAL RECALL Serial numbers 109017 thru 115660. EXPANDED RECALL serial numbers104205 thru 117464. Model 1175: INITIAL RECALL serial numbers 200731 thru 202199. EXPANDED RECALL serial numbers 200251 thru 202787 Model 1273: INITIAL RECALL serial numbers 315516 thru 324528. EXPANDED RECALL serial numbers 306216 thru 601338. Model 1274: INITIAL RECALL serial numbers 296080 thru 496546. EXPANDED RECALL serial numbers 404822 thru 498144. Model 1275: INITIAL RECALL serial numbers 500705 thru 501661. EXPANDED RECALL serial numbers 500211 thru 502015.
510(k) numbers
not on file
PMA numbers
["P940031"]
Product code
LWP
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register