Skip to the content

FDA Medical Device Recalls (openFDA)

Orthosensor, Inc.: Verasense Knee System, Model SYK-TRCR 02. Intelligent disposable tibial insert that measures dynamic loads during orthopedic knee surgery.

Recall number
Z-1029-2015
Recalling firm
Orthosensor, Inc.
Product
Verasense Knee System, Model SYK-TRCR 02. Intelligent disposable tibial insert that measures dynamic loads during orthopedic knee surgery.
Status
Terminated
Initiated
2014-12-05
Posted
2015-01-29
Terminated
2015-03-13
Reason
Presence of one inoperable (dead) sensor pad.
Root cause
Under Investigation by firm
Action
OrthoSensor sent an Recall letter dated December 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Required Action for Users " Remove the above-referenced serial numbers from your VERASENSE inventory and quarantine. " Complete the enclosed Customer Response Form. Actions to be taken by OrthoSensor " You will be credited for these devices for your full purchase amount. The Returned Materials Authorization (RMA) number for each device in your possession is listed above. " OrthoSensor will replenish these devices as soon as they are available. For further questions please call (954) 577-7770.
Quantity
22 devices
Distribution
US Distribution to the states of : IL, MI, NY, CA, VA and NV.
Lot, serial or model codes
Model SYK-TRCR 02, Lot # 091914V11770, Serial #s: 411654008, 411585023, 411654007, 411585032, 411585014, 411654016, 411585017, 411654015, 411585030, 411654020, 411654004, 411654010, 411585007, 411585005, 411654011, 411585036, 411585035, 411654018, 411654012, 411585011, 411654014, 411654005, 411654001, 411585003.
510(k) numbers
K130380
PMA numbers
not on file
Product code
ONN
Firm FEI
3008592715
Firm city
Sunrise
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register