Impac Medical Systems Inc: SEQUENCER component of MOSAIQ versions 1.30 (all builds), 1.40 (all builds), 1.50 (all builds), 1.60 (all builds), and 2.0 (all builds). Manufactured by IMPAC Medical Systems, Inc. Sunnyvale, CA. The intended use of SEQUENCER (a verification system) is to assist in the process of patient treatment on a radiotherapy treatment machine. SEQUENCER provides the capability to notify the users of actions that need to take place prior to treatment, to display reference images for set up purposes, and to auto setup the machine to predefined settings. Actual settings are read from the treatment machine, via the machine's communication interface, and compared to the predefined values. If a mismatch occurs between the planned values and the actual machine settings, treatment is prohibited. At appropriate times during the treatment, the actual delivered values are recorded to provide treatment tracking. Treatments are initiated by the operator, via the machine control system, only after the user has verified that the treatment setup is correct. SEQUENCER does not bypass any checks put in place by the machine's control system.

Recall number
Z-1027-2011
Recalling firm
Impac Medical Systems Inc
Product
SEQUENCER component of MOSAIQ versions 1.30 (all builds), 1.40 (all builds), 1.50 (all builds), 1.60 (all builds), and 2.0 (all builds). Manufactured by IMPAC Medical Systems, Inc. Sunnyvale, CA. The intended use of SEQUENCER (a verification system) is to assist in the process of patient treatment on a radiotherapy treatment machine. SEQUENCER provides the capability to notify the users of actions that need to take place prior to treatment, to display reference images for set up purposes, and to auto setup the machine to predefined settings. Actual settings are read from the treatment machine, via the machine's communication interface, and compared to the predefined values. If a mismatch occurs between the planned values and the actual machine settings, treatment is prohibited. At appropriate times during the treatment, the actual delivered values are recorded to provide treatment tracking. Treatments are initiated by the operator, via the machine control system, only after the user has verified that the treatment setup is correct. SEQUENCER does not bypass any checks put in place by the machine's control system.
Status
Terminated
Initiated
2009-07-17
Posted
2009-08-21
Terminated
2011-11-25
Reason
Potential for improper treatment. A software error may result in potential major underdosing to targeted areas from jaw blockage of the MLC aperture (jaw intrusion), and dosing to non-targeted areas from interstitial leaf leakage (jaw extrusion).
Root cause
Software design
Action
Impac Medical Systems Inc sent an Important Safety Notice letter to all affected consginees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact Impact Software support as soon as possible if they have the affected configurations identified in the letter. Customers were asked to complete the form included with the letter and and fax to: European customers: +44 1293 654401 American & rest of world customers 702-992-5002 or e-mail to support@impac.com Customers were asked to distribute the Safety Notice to any and all users of IMPAC software at their organization who are potentially affected by this issue. For questions e-mail support@impac.com or call European customers: 44 1293 654320 American & rest of world customers: 800-488-4672.
Quantity
not on file
Distribution
Worldwide Distribution - USA including AL, CA, CT, FL, GA, IA, IL, KY, MI, MN, MS, ND, NJ, OH, OR, PA,TN, TX, VA, WA, WI, and WV and the country of Canada.
Lot, serial or model codes
MOSAIQ version 1.30 (all builds), MOSAIQ version 1.40 (all builds), MOSAIQ version 1.50 (all builds), MOSAIQ version 1.60 (all builds, and MOSAIQ version 2.0 (all builds).
510(k) numbers
K981313
PMA numbers
not on file
Product code
IYE
Firm FEI
1000123805
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source