Philips Electronics North America Corporation: SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system

Recall number
Z-1024-2016
Recalling firm
Philips Electronics North America Corporation
Product
SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system
Status
Terminated
Initiated
2015-07-06
Posted
2016-02-26
Terminated
2017-08-29
Reason
Wrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to instructions for use can lead to patient burns.
Root cause
Device Design
Action
On 6/18/2015, the Field Safety Notice / Medical Device Correction Notice was sent to customers via certified mailing. The letter explains the issue and provides user instructions. A sleeve is being developed that will be applied to the cable on the Sense Body Coil 1.5T and provide consistent spacing between the cable and the patient, thereby reducing the chance of RF interaction and heating. This sleeve will be distributed as part of a Field Change Order with reference FCO781 00437 for Sense Body Coils in the installed base and will be installed by the Philips Field Service Engineers.
Quantity
5672
Distribution
US: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jersey Jordan Kazakhstan Kenya Korea, Republic of Kuwait Kyrgyzstan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Malta Martinique Mauritius Mexico Monaco Mongolia Morocco Mozambique Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian territory, occupied Panama Papua New Guinea Peru Philippines Poland Portugal Puerto Rico Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Venezuela Viet Nam Yemen Zimbabwe
Lot, serial or model codes
4522 131 5575x 4522 132 1985x 4522 132 7619x 4598 000 5187x 4598 000 5189x 4598 003 59051
510(k) numbers
K001796, K013344, K030520, K945945
PMA numbers
not on file
Product code
LNH
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source