Stryker Spine: Stryker Trio Small Offset Connector Non Sterile; Manufactured by Stryker Spine SAS, Z.I. Marticot -33610- Cestas, France. Distributed in USA by Howmedica Osteonics Corp., Mahwah, NJ. The Trio Spinal Fixation Systems are pedicle screws systems intended for use in providing immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar, and sacral spine: degenerative spondyolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion.
- Recall number
- Z-0986-2009
- Recalling firm
- Stryker Spine
- Product
- Stryker Trio Small Offset Connector Non Sterile; Manufactured by Stryker Spine SAS, Z.I. Marticot -33610- Cestas, France. Distributed in USA by Howmedica Osteonics Corp., Mahwah, NJ. The Trio Spinal Fixation Systems are pedicle screws systems intended for use in providing immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar, and sacral spine: degenerative spondyolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion.
- Status
- Terminated
- Initiated
- 2006-09-05
- Posted
- 2009-04-10
- Terminated
- 2009-04-14
- Reason
- Reports were received of the ball ring ejecting from the connector's body when other connectors have already been locked down.
- Root cause
- Device Design
- Action
- A notification letter dated September 12, 2006 and a Product Accountability Form were sent via Federal Express (with return receipt) to all affected branches/agencies. Further information can be found by calling 1-201-760-8256.
- Quantity
- 2,417 units
- Distribution
- Nationwide Distribution including OR, CA, MA, VA, GA, FL, NY, NM, NJ, UT, KS, OH, MI, MN, IN, MT, TX, WI, AL, CO, and TN.
- Lot, serial or model codes
- Catalog Number 48902015; Lot codes: 056254, 056255, 056256, 056257, 56258, 056259, 058065, 05A140, 05A142, 058066, 05A139, 05A141, 058915, 05A143, 05A145, 05A146, 05A144, 061075, 061077, 061551, 061072, 061076, 0615550, 061549, 061760, 05A148, 05A149, and 05A147.
- 510(k) numbers
- K052971
- PMA numbers
- not on file
- Product code
- MNI
- Firm FEI
- 3004024955
- Firm city
- Allendale
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28