FDA Medical Device Recalls (openFDA)

Draegar Medical Systems, Inc.: Omega Systems These devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

Recall number
Z-0975-2019
Recalling firm
Draegar Medical Systems, Inc.
Product
Omega Systems These devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
Status
Open, Classified
Initiated
2019-02-08
Posted
not on file
Terminated
not on file
Reason
The firm is recalling their Delta family of patient monitors software due to cybersecurity vulnerabilities, which may cause the device to reboot, lose functionality, and/or lose communication.
Root cause
Software design
Action
On February 8, 2019, the firm, Drager, notified affected customers of the recall via an "Urgent: Medical Device Recall" letter. Customers were informed that the existing software had cybersecurity vulnerabilities and that software version VF10.1 would mitigate these issues. Customers were asked to complete a Customer Acknowledgment and Response Form and return it to Drager via fax 215-372-2940 or email to ditelford.guality@draeger.com. Your local Draeger Service representative will contact you to schedule an appointment for upgrading your systems software once a compatible version has been released for distribution. If you are aware of any incidents related to this issue or if you have any further questions regarding operation of your patient monitor, please contact Drager Service Technical Support between the hours of 8:00 AM- 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2 again).
Quantity
10699 in total
Distribution
US Nationwide distribution including Puerto Rico.
Lot, serial or model codes
All software versions, all serial numbers.
510(k) numbers
["K141756"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3005783425
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register