FDA Medical Device Recalls (openFDA)
Intuitive Surgical, Inc.: da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
- Recall number
- Z-0964-2026
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
- Status
- Open, Classified
- Initiated
- 2024-08-21
- Posted
- 2025-12-19
- Terminated
- not on file
- Reason
- Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
- Root cause
- Device Design
- Action
- On 8/21/2024, correction notices were sent to customers who were informed of the following: 1) Share this notification with all personnel at your site who perform and support da Vinci procedures. 2) Inform affected personnel when the correction has been completed. 3) Complete and return the acknowledgement form via email to Recalls@intusurg.com 4) Retain a copy of this letter and the acknowledgement form for your files. 5) Firm is actively working on making updated motor module assemblies available. Once available, an Intuitive representative will schedule a visit to replace the affected motor module sub-assembly. If you need further information or support concerning this notification, please contact the firm's Customer Service: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com
- Quantity
- 144
- Distribution
- US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.
- Lot, serial or model codes
- UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021, 10576022, 10576023, 10576024, 10576025, 10576026, 10576027, 10576028, 10579442, 10579446, 10579453, 10579455, 10579456, 10579457, 10579461, 10579466, 10582427, 10582428, 10582429, 10591177, 10591179, 10591180, 10591184, 10591197, 10594614, 10594615, 10594616, 10594617, 10594618, 10594619, 10597107, 10597108, 10597109, 10597110, 10599042, 10599044, 10599045, 10599048, 10600750, 10600753, 10603334, 10603335, 10603343, 10603344, 10603345, 10603367, 10605468, 10606129, 10606133
- 510(k) numbers
- ["K232610"]
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28