Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Biomet Oxford Unicompartmental Knee, Phase 3, Right Medical Tibial Tray, Standard Size B, CoCr, sterile, Biomet UK Ltd, Bridgend, UK; . Unicompartmental knee joint replacement system. Designed to allow for motion between the plastic tibial bearing and metal tibial baseplate components of a unicompartmental knee prosthesis.

Recall number
Z-0962-2009
Recalling firm
Biomet, Inc.
Product
Biomet Oxford Unicompartmental Knee, Phase 3, Right Medical Tibial Tray, Standard Size B, CoCr, sterile, Biomet UK Ltd, Bridgend, UK; . Unicompartmental knee joint replacement system. Designed to allow for motion between the plastic tibial bearing and metal tibial baseplate components of a unicompartmental knee prosthesis.
Status
Terminated
Initiated
2008-12-19
Posted
2009-02-03
Terminated
2009-07-07
Reason
Packages labeled as "right" actually contain a "left" component.
Root cause
Packaging process control
Action
Biomet salesmen were notified by phone and instructed to retrieve the products. One customer was notified by letter dated 12/19/08 that use of the product should cease and that a salesman would pick up the product. A second customer had implanted the product and was hand delivered a letter dated 12/19/08 instructing that the patient should be monitored and symptoms treated as necessary. Contact Biomet at 574-362-3983 for assistance.
Quantity
2
Distribution
Massachusetts and West Virginia.
Lot, serial or model codes
Lot 1563796.
510(k) numbers
not on file
PMA numbers
["P010014"]
Product code
NRA
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register