Ion Beam Applications S.A.: Proteus 235, Pencil Beam Scanning, PTS-9.4.X, PTS-10.0.1.1 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Recall number
Z-0943-2018
Recalling firm
Ion Beam Applications S.A.
Product
Proteus 235, Pencil Beam Scanning, PTS-9.4.X, PTS-10.0.1.1 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
Status
Terminated
Initiated
2017-11-01
Posted
not on file
Terminated
2019-03-29
Reason
The internal configuration of the electrometers and the Real-Time control boards of the Dekimo Scanning Controller is not checked before each patient treatment.
Root cause
Device Design
Action
The firm, IBA, sent an "Urgent Field Safety Notice" dated 10/31/2017 to customers on 10/31/2017. The notice described the product, problem and actions to be taken. stating the following: There is no specific recommendation for the user. IBA will provide an Internal User notice to the IBA site team, requesting the operator to select the Beam Delivery Point CGTR, by clicking on the BDP button, before each patient treatment preparation (see Annex 1 for details). This will begin the check of the internal configuration of all the elements of the Dekimo Scanning Controller. An updated system will be deployed on February 28, 2017. If you have any questions, contact Customer Complaints & Vigilance Director at Vigilance@iba-group.com or +32 10 203 787 or helpdesk at +32 2 507 20 81 (available 24/7).
Quantity
3
Distribution
Worldwide Distribution: US (nationwide) to: MI only and country of: France.
Lot, serial or model codes
SBF.101 (EU), SBF.105 (US)
510(k) numbers
["K152224"]
PMA numbers
not on file
Product code
LHN
Firm FEI
3000256071
Firm city
Louvain La Neuve
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register