Intuitive Surgical, Inc.: IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000. Product number 371683-05/Model Number VS3000 The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and noncardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.

Recall number
Z-0940-2015
Recalling firm
Intuitive Surgical, Inc.
Product
IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000. Product number 371683-05/Model Number VS3000 The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and noncardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.
Status
Terminated
Initiated
2014-12-05
Posted
2015-01-07
Terminated
2015-03-03
Reason
Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wheel bolts gradually unthreading and the wheels loosening.
Root cause
Device Design
Action
The firm, Intuitive Surgical, sent an "Urgent Medical Device Correction" letter dated December 5, 2014 to all customers on December 5, 2014 by trackable method. The letter described the product, problem and actions to be taken. The customers were instructed to: ensure all affected personnel are fully informed of this notice; identify affected product, and retain a copy of this notice with your da Vinci Si user manual. Please note: the da Vinci Si Vision Side Cart can continue to be used safely. Ensure that all wheels are fully unlocked before attempting to move the Vision Side Cart. Avoid rolling the Vision Side Cart over uneven surfaces (i.e., over cables or a threshold). Intuitive Surgical is working on a newly designed wheel assembly. Upon availability, Field Service will retrofit the wheel assembly of the affected Vision Side Carts with the new design. If you need further information or support concerning this Medical Device Correction, please contact your Clinical Sales Representative or Intuitive Surgical Customer Service at the numbers listed below: North America and South America: 8008761310 Option 3 (6 a.m. to 5 p.m. PST) Japan: 0120565635 or 0355751362 (9 a.m. to 6 p.m. JST) South Korea: 0232713200 (9 a.m. to 6 p.m. KST) Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 a.m. to 6 p.m.) or ics@intusurg.com
Quantity
202
Distribution
Worldwide Distribution: US (nationwide) and countries of: Belgium, France, Germany, Romania, India, Israel, Italy, South Korea, Norway, Slovak Republic, Spain, Sweden, Turkey, and United Kingdom.
Lot, serial or model codes
Product number 371683-05/Model Number VS3000 Serial numbers - USA: SH0018, SH0057, SH0097, SH0132, SH0165, SH0235, SH0019, SH0058, SH0098, SH0133, SH0167, SH0248, SH0021, SH0059, SH0099, SH0134, SH0168, USG015, SH0024, SH0060, SH0100, SH0135, SH0169, USG072, SH0025,SH0064, SH0101, SH0136, SH0170, USG146, SH0026, SH0065, SH0102, SH0137,SH0171, USG213, SH0029, SH0066, SH0103, SH0138, SH0172, USG317, SH0030, SH0067, SH0104, SH0139, SH0175, USG352, SH0031, SH0068, SH0105, SH0140, SH0177, USG353, SH0032, SH0069, SH0106 ,SH0141, SH0179 ,USG445, SH0033, SH0070, SH0107, SH0142, SH0180, USG519, SH0035, SH0071, SH0109, SH0143, SH0182, USG559, SH0036, SH0072, SH0110, SH0144, SH0184, USG631,SH0038, SH0073, SH0112, SH0145, SH0185, USG700, SH0039, SH0076, SH0113, SH0146, SH0186, USG701,SH0040, SH0077, SH0115, SH0147, SH0187, USG702, SH0041, SH0078, SH0117, SH0149, SH0188 ,USG703, SH0042, SH0079, SH0118, SH0150, SH0189, USG721, SH0044, SH0080, SH0119, SH0151, SH0190 ,USG728, SH0045, SH0081, SH0120, SH0152, SH0194, USG748, SH0046, SH0082, SH0121, SH0153, SH0196, USG750, SH0047, SH0085, SH0123, SH0154, SH020,2 USG764, SH0048, SH0088, SH0124, SH0155, SH0203, USG769, SH0050, SH0089, SH0125, SH0156, SH0204, USG770, SH0051 SH0090 SH0126 SH0157 SH0206 USG775, SH0052, SH0091, SH0127, SH0158, SH0211, USG786, SH0053, SH0092, SH0128, SH0159, SH0215,, USG787, SH0054, SH0094, SH0129, SH0160, SH0216, USG788, SH0055, SH0095, SH0130, SH0163, SH0223, SH0056, SH0096, SH0131, SH0164, SH0224, Out of US: SH0034,USG217, SH0086, SH0087, SH0028,SH0084,SH0111, SH0191, USG754, SH0114, SH0116, SH0063, SH0148, USG484, USG727, SH0161, SH0162, SH0061, SH0093, SH0238, USG726, SH0062, SH0037, USG806.
510(k) numbers
["K081137"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register