FDA Medical Device Recalls (openFDA)

Philips North America, LLC: The IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for the following product model #s: 21075A, 21076A, 21078A, 989803162621, 989803162631, 989803162641.

Recall number
Z-0926-2022
Recalling firm
Philips North America, LLC
Product
The IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for the following product model #s: 21075A, 21076A, 21078A, 989803162621, 989803162631, 989803162641.
Status
Open, Classified
Initiated
2022-02-07
Posted
not on file
Terminated
not on file
Reason
Update to instructions for use regarding the cleaning and disinfection process for the reusable probes.
Root cause
Device Design
Action
The firm, Philips, issued "URGENT Medical Device Recall" Customer notification letters on 07 Feb 2022 via certified mail. Actions planned by the firm: An IFU addendum (part# 453564978601) was created that provides updated instructions for cleaning, disinfection, and sterilization. This addendum is to be downloaded www.philips.com/ifu and kept with the main IFU (part# 453564635891). Customer should take the following actions to prevent risk to patients or users: Prepare and soak in an enzymatic solution (detergent) to remove any debris prior to disinfection or sterilization. To reduce risk of cross-infection always clean all probes before applying to a different patient. Esophageal/Rectal Probes must be cleaned and sterilized between patient use. Skin probes can be cleaned and disinfected or cleaned and sterilized between patient use. Contact Philips Customer Services with questions. US customers: 1 (800) 722-9377, Worldwide customers: Contact local Philips representative.
Quantity
478,931 units
Distribution
Worldwide distribution: US (nationwide) Including Puerto Rico and OUS (foreign) countries including: Afghanistan, Albania, Algeria, Angola, Argentina, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burundi, Cambodia, Canada, Chile, China, Columbia, Congo (Democratic Republic of the), Costa Rica, Cote D'lvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, French Guiana, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Japan, Jordan, Kazakhstan, Korea Republic of, Kuwait, Lao People's Democratic Republic, Latvia, Lebanon, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives, Mali, Malta, Mauritius, Mayotte, Mexico, Moldova, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saint Lucia, Saint Pierre and Miquelon, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Swaziland, Sweden, Switzerland, Syrian and Republic, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen and Zambia.
Lot, serial or model codes
All lots manufactured IFUs from 2008 to present. UDI for probe model 21075A (01)00884838000995, model 21076A (01)00884838001008, model 21078A (01)00884838001015, model 989803162621 (01)00884838008045, model 989803162631 (01)00884838008052, and model 989803162641 (01) 00884838008069
510(k) numbers
["K962070"]
PMA numbers
not on file
Product code
FFL
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register