Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc: Medtronic Orthopedic Dovetail Clamp, REF 9730289, Medtronic Navigation, Louisville, CO 80027. In combination with a tera-tracker reference frame, is used by an Orthopedic surgeon to navigate the location of an acetabular cup positioner/impactor during hip replacement surgery.

Recall number
Z-0923-2010
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Orthopedic Dovetail Clamp, REF 9730289, Medtronic Navigation, Louisville, CO 80027. In combination with a tera-tracker reference frame, is used by an Orthopedic surgeon to navigate the location of an acetabular cup positioner/impactor during hip replacement surgery.
Status
Terminated
Initiated
2009-11-03
Posted
2010-03-03
Terminated
2010-08-13
Reason
Orthopedic clamp may slip if used with non-indicated system, resulting in possible inaccurate navigation information being displayed.
Root cause
Error in labeling
Action
All affected users were notified by a Product Correction Notification letter on 11/03/2009 providing a description of the issue and informing them of the updated instructions for use for the Orthopedic Dovetail Clamp. Customers are to ensure that all users read the letter, place a copy of the letter with the IFUs for the related software applications listed in the letter, and fill out the attached response form indicating the receipt of the information. Further information is available at 720-890-3309.
Quantity
244 units
Distribution
Worldwide Distribution -- USA, including VA, IA, MN and OK and countries of Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Hong Kong, Iceland, India, Israel, Italy, Japan, Korea, Lebanon, Malta, Mexico, Netherlands, Poland, Russian Federation, Saudi Arabia, South Africa, Switzerland, Turkey, United Arab Emirates, and UK.
Lot, serial or model codes
All lot numbers.
510(k) numbers
["K021980"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register