Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: iQ200 Series Urine Microscopy Analyzer, Catalog numbers: 700-3345, 700-3375, 700-3325, 700-3347, 700-3300RFB, 700- 3320, 700-3320RFB, 700-3322, 700-3322RFB, 700-3326, 700- 3340, 700-3370, B48999, C10683, C10684 - Product Usage: device used to automate the complete urinalysis profile, including urine test strip chemistry panel and microscopic sediment analysis.

Recall number
Z-0914-2020
Recalling firm
Beckman Coulter Inc.
Product
iQ200 Series Urine Microscopy Analyzer, Catalog numbers: 700-3345, 700-3375, 700-3325, 700-3347, 700-3300RFB, 700- 3320, 700-3320RFB, 700-3322, 700-3322RFB, 700-3326, 700- 3340, 700-3370, B48999, C10683, C10684 - Product Usage: device used to automate the complete urinalysis profile, including urine test strip chemistry panel and microscopic sediment analysis.
Status
Terminated
Initiated
2019-10-28
Posted
not on file
Terminated
2026-04-01
Reason
Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).
Root cause
Software in the Use Environment
Action
An Urgent Medical Device Recall letter was sent during the week of 10/28/2019 to the affected customers via mail and e-mail. The notice stated the misidentification issue as follows: Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics)." The following actions were provided to reduce the potential for this issue until a software update is available: - Follow good laboratory practice (GLP) and ensure that a unique identification system is in place. - Ensure that specimens are labeled with a unique Specimen ID and manual orders utilize a unique Specimen ID. - Do not reuse barcode number that identify different specimens. If barcode numbers must be reused, ensure results have been released from the LIS and the instrument Work List has been cleared. - Clear all manual Orders immediately after results are released, even if barcodes are used for subsequent runs (see iQ200 or iChemVELOCITY IFU, Chaper 7, Manual Orders).
Quantity
6601 devices
Distribution
Worldwide distributions.
Lot, serial or model codes
UDI: 10837461001317, 10837461001355, 10837461001256,10837461001300; All Part Numbers
510(k) numbers
K022774
PMA numbers
not on file
Product code
LKM
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source