Skip to the content

FDA Medical Device Recalls (openFDA)

EP Medsystems: EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator. The stimulator is capable of single channel, simultaneous and sequential stimulation at programmed output settings using built-in protocols, and includes an emergency stimulation function.

Recall number
Z-0914-06
Recalling firm
EP Medsystems
Product
EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator. The stimulator is capable of single channel, simultaneous and sequential stimulation at programmed output settings using built-in protocols, and includes an emergency stimulation function.
Status
Terminated
Initiated
2006-04-04
Posted
2006-05-31
Terminated
2007-04-24
Reason
Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation unit-2)-Presence of a DC offset voltage on an output channel-3)-Inability to synchronize with a non-EP MedSystems recording system
Root cause
Other
Action
The firm initiated the recall on April 4, 2006 through a notification letter sent to the electrophysiology lab managers at the healthcare institutions and two domestic distributors to which the device has been distributed. The firm used a traceable courier service to deliver notification of the recall.
Quantity
187 units
Distribution
Product was shipped to hospitals/users throughout the US. The firm has a Veterans Affair contract number V797P-4650A. Also the firm has shipped the devices to Germany, United Kingdom, Netherlands, Italy, Japan, France, Saudi Arabia, Hong Kong, Syria, Malaysia, Sweden, Armenia, Australia, Canada, Iceland, S. Africa, and Taiwan.
Lot, serial or model codes
Model number: EP-4-02 and EP-4-04.  EP40312-001 through 010; EP40403-001 through 014; EP40406-001 through 042; EP40408-001 through 039; EP40410-001 through 040; EP40411-001 thourgh 040; EP40504-001 through 040; EP40506-001 through 040; EP40508-001 through 020; EP40510-001 through 020; EP40511-001 through 040.
510(k) numbers
["K040207","K041442"]
PMA numbers
not on file
Product code
JOQ
Firm FEI
2126673
Firm city
West Berlin
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register